Operating Instructions Orthophos SL 2D / 3D
5.1 Acquiring the X-ray image
5.1.2.1.3
Selecting an exposure program
✔
The unit is switched on and ready for exposure.
1. Touch the PAN symbol at the top of the touchscreen.
ª
The program group PAN is selected.
2. Select an exposure program. To do this, touch the arrow keys + C
and -A. If a subroutine, e.g. without artifacts or 1.25x magnification, is
available for this program, the program display is grayed out. Touch
the program display B several times. All subroutines of the selected
program are displayed one by one.
3. Follow the instructions in the comment line on the touchscreen. If
required, press the R key to move the unit back to the starting
position.
ª
The diaphragm and the sensor move into the starting position.
ª
The exposure program is selected.
CAUTION
Pressing the R key moves the unit to the starting position.
A patient positioned in the unit may be injured by moving parts.
➢ Check that a patient is not positioned in the unit before moving it to
the starting position.
C
B
A
Ready for exposure
NOTE
The PAN/3D sensor unit is rotated via a motor drive.
The gearing of the sensor unit can be damaged if it is turned by hand.
➢ Press the R key to rotate the sensor unit back to the starting position
via the motor drive. The sensor unit is always rotated in combination
with the entire main rotation unit. The combisensor travels to the
proper starting position for panoramic, Ceph, or 3D exposures,
depending on the program group selected.