119
Dentsply Sirona
5 Operation
Operating Instructions Orthophos SL 2D / 3D
5.1 Acquiring the X-ray image
XGPF-36246 / XGPF-36248 / XGPF-38075; Patient information clamping
✔
The program settings have been made.
✔
The patient is positioned in the unit.
✔
No further auxiliary messages may be displayed in the comment line
of the touchscreen. The
"Ready for exposure"
message must
appear.
Tip: If you press the exposure release button on the remote control
while the door is open, the
"Close the door"
message is displayed
with the help code H321. Close the door and acknowledge the
message.
1. Press release button (A) and hold it down until the end of the
exposure.
ª
The exposure is released.
"Exposure is performed"
appears in
the comment line on the touchscreen. During radiation, the
optical radiation indicator (B) lights up on the Easypad. In
addition, an acoustic signal sounds throughout the radiation.
Radiation can be released several times during the exposure.
2. Press and hold release button A. Wait until a short pulsed tone
sequence sounds (can be deactivated by a service engineer). The
message
"Please wait"
appears in the comment line of the
touchscreen followed by confirmation of the exposure data. Exposure
mode, exposure program, tube voltage and current, actual radiation
time and dose-area product are displayed.
ª
The forehead and temple or temporomandibular joint supports
open automatically.
3. Let go of release button (A).
ª
The exposure is completed.
ª
The X-ray image is displayed on the PC monitor after a brief
period.
4. Guide the patient out of the unit.
IMPORTANT
Premature release of a new exposure is prevented by the automatic
exposure blocking function. This function is used for thermal protection
of the X-ray tubes.
After the release button is pressed, the message
"Ready for exposure
in "XX" seconds"
appears in the comment line of the touchscreen. The
remaining cooling time is counted down and is displayed under "XX".
Only after the cooling period has elapsed is it possible to release a new
exposure.
IMPORTANT
Advise the patient of what to do during the exposure and check to make
sure that they comply:
- The patient must not move his/her head in any way.
- The patient's shoulders must not be hunched.
- For cephalometric images, the arms of the patient must hang down
freely at the sides.