6 Service
Dentsply Sirona
6.1 Cleaning and care
Operating Instructions Orthophos SL 2D / 3D
6.1.4
Cleaning the occlusal bite block
If the hinges of the occlusal bite block begin to squeak during operation
after prolonged periods of use, they must be cleaned.
1. Unplug the occlusal bite block from its holder on the unit.
2. Gently push apart the guide mandrel of the lever (A) on the bite block
plate and the eyelet of the rod (B) in the direction of the arrows and
unhinge the lever.
3. Swivel the bite block plate (C) vertically upward so that the lever (A)
is pointing downwards.
4. Pull the bite block plate (C) forwards out of its hinge.
5. Clean the hinge axles (D) and the guide lugs (E) with disinfectant.
6. Assemble the occlusal bite block by following the same procedure in
reverse order. When assembling, check the position of the bite block
plate; the segment must point toward the connecting lever.
7. Plug the occlusal bite block back into its holder on the unit.
B
A
C
A
C
D
E