Fig u re 37 - Switch DUT to "ON" for upcoming SF:L2O tests.
Note the purpose of the "STOP BEFORE..." break is to allow you to accommodate power-sensitive
equipment, such as a personal computer or equivalent DUT, which cannot be arbitrarily
disconnected from the AC supply without potentially damaging the DUT. This type of equipment is
much less common than equipment that can be switched "OFF" at any time. The more typical
setting of "STOP BEFORE..." is therefore "NO", in which case the BREAK menus of
Fig u re 36
and
Fig u re 37
will not be displayed.
Many types of medical equipment do not fully disconnect from the AC supply when the power
switch is set to "OFF". Even when "OFF", this type of equipment remains effectively connected to
the AC supply and continues to operate in a standby state, for example, by charging an internal
battery. For this type of equipment, "DUT OFF" leakage tests are required under IEC60601 and
AAMI-ES1 standards.
"DUT OFF" leakage measurements are optionally enabled by the user before the test is performed,
by means of the leakage setup menu. If "DUT OFF" tests are enabled, ES601
Plus
will stop after all
leakage measurements have been completed with DUT power switch in "ON" position. ES601
Plus
will then display instructions as shown in the following figure.
Manual Testing/Chapter 5
#
Page 78
Содержание ES601 Plus
Страница 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Страница 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Страница 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Страница 59: ...ES601Plus OPERATING MANUAL Fig ure 13 Connecting to a line powered DUT Connecting To ES601 Chapter 4 Page 40 ...
Страница 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Страница 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Страница 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Страница 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Страница 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Страница 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Страница 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Страница 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Страница 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...