OPERATING MANUAL
8.5.22 Protocol (VDE0751 and IEC62353)
The Protocol defines the device and patient leakage measurement protocols under VDE0751 and
IEC62353 safety standards. In the VDE Alternate Method, the DUT is connected to an isolated
mains source rather than the AC supply. In the Direct Method, the DUT is powered by the AC
supply, and patient leakage is determined by connecting isolated applied parts to mains only (Type B
parts are not tested). For VDE0751 and IEC62353 standards, the "IEC60601 method" means the
DUT is powered by the AC supply and patient leakage is determined by connecting applied parts
either to mains or to ground (Type B parts will be tested).
The Alternate method of leakage measurement can only be used with Type F
Isolated Leads.
8.5.23 Test Point-to-Point (IEC61010)
Clicking this check box enables measurement of Enclosure Leakage Current from one arbitrary
point on the DUT enclosure to a second arbitrary point. The point-to-point enclosure connections
are made via jacks AP1 and AP2 on ES601
Plus
. This test is pertinent to the IEC61010 standard
only.
8.5.24 Uploading AutoSequences
To upload Safety AutoSequences to ES601
Plus
, press the “Upload To ES601" button. Choose
either uploading via USB, Ethernet, or Serial Port, and then press "Connect". If the connection is
via ethernet, enter the IP address assigned to ES601
Plus
(the IP address of ES601
Plus
is displayed
on the LCD when the analyzer is operating in Ethernet Mode). If the connection is via serial port,
choose the COM port to which the ES601
Plus
is connected.
After pressing "Connect", a message box with a transfer progress bar will appear. After the transfer
is completed, a Transfer Complete message will appear and the message box will disappear.
8.5.25 Downloading AutoSequences
Sometimes it is desirable to download sequences stored in ES601
Plus
. You can do this by pressing
the "Download From ES601" button. Choose to download via USB, Ethernet, or Serial Port. Doing
so will overwrite the configuration of all 24 Safety AutoSequences in ES601pc with the sequences
stored in ES601
Plus
.
These sequences can the be changed and uploaded back to ES601
Plus
.
ES601pc Software/Chapter 8
#
Page 152
Содержание ES601 Plus
Страница 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Страница 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Страница 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Страница 59: ...ES601Plus OPERATING MANUAL Fig ure 13 Connecting to a line powered DUT Connecting To ES601 Chapter 4 Page 40 ...
Страница 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Страница 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Страница 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Страница 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Страница 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Страница 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Страница 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Страница 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Страница 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...