CALIBRATE
OFFSET:
The "
OFFSET:
" displayed to the right of this control is the resistance of
the test lead used in measuring point-to-point resistance. In a point-to-
point test, the
OFFSET
calibration is subtracted from the measured
resistance to obtain the result which is displayed by ES601
Plus
.
Calibrate the
OFFSET
when required, particularly if the test lead used for
point-to-point measurement is changed. Performing the calibration from
System Settings Menu #2 (see Section 3.3) records the
OFFSET
in non-
volatile memory. Performing the calibration from the Sequence Controls
menu updates the
OFFSET
for the manual test, but does not store the
calibration in the non-volatile memory.
Calibrate the
OFFSET
according to this procedure:
1.
Connect the Kelvin Cable to the corresponding jacks of
ES601
Plus
.
2.
Connect the test lead which will be used for point-to-point
measurement to the "TO DUT CASE" jack of ES601
Plus
.
3.
Connect the free ends of the Kelvin Cable and the test lead
together. Verify a solid mechanical contact exists between the
two cables at the far end.
4.
Press the
CALIBRATE OFFSET
button. The resistance will
be measured and displayed within one second.
5.
The result should be 0.05 ohms or lower for a typical test lead of
2-meter length. For best results, avoid use of a poor-quality or
excessively long cable having a resistance exceeding 0.1 ohm.
ES601
Plus
will produce an alarm and reject the calibration if the
resistance exceeds 3 ohms.
6.
If the resistance seems unreasonably high, verify the connection
between the test lead and the Kelvin Cable. Also confirm the
test lead makes a solid contact when inserted into the "TO DUT
CASE" jack. Repeat the calibration as required until an
acceptable result is obtained.
ESCAPE:
This button on the Sequence Controls menu aborts the test in progress
and returns the display to the MANUAL MODE menu.
Manual Testing/Chapter 5
#
Page 58
Содержание ES601 Plus
Страница 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Страница 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Страница 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Страница 59: ...ES601Plus OPERATING MANUAL Fig ure 13 Connecting to a line powered DUT Connecting To ES601 Chapter 4 Page 40 ...
Страница 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Страница 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Страница 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Страница 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Страница 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Страница 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Страница 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Страница 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Страница 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...