6.2 Automated Testing Principles
ES601
Plus
provides automated testing through General Inspections, Tester AutoSequences, and
Test Procedures. Each type of automated test is based on a sequential list of
In s tru c tio n s
that an
operator can step through using ES601
Plus
touch screen interface.
6.2.1 ES601
Plus
Instructions
ES601
Plus
Instructions comprise a
c o m m an d
and an
o p c o d e .
The op code defines what the
Instruction does and what buttons are shown on the touch screen of ES601
Plus
when the
Instruction is executed. Depending on the op code, the user may be required to enter a
measurement value or a comment via the QWERTY touch keypad when that Instruction is
encountered in the sequence. Some Instruction op codes cause the command to be transmitted to
an external device via a selected
T e s te r Po rt
, these RS232 serial ports being located on the left side
panel of ES601
Plus
.
6.2.2 General Inspections
ES601
Plus
provides 18 General Inspections, which are user-programmable by means of the
AutoSequence Editor of the ES601pc software. A General Inspection can comprise from 1 to 24
steps, where each step corresponds to an ES601
Plus
Instruction. In principle, General Inspections
are intended for work tasks in which the user will be visually inspecting the equipment under test,
and will be recording simple "pass" or "fail" results for each item inspected. General Inspections are
intended to be fairly generic and are targeted at an overall class of medical equipment, such as "IV
Pumps" or "Defibrillators". Such Inspections might include steps such as "CHECK PUMP
ROLLERS FOR WEAR", or "CHECK DEFIB PADDLES FOR PITTING/SCRATCHES".
General Inspections need not be limited to simple "pass/fail" visual tests, however. Through the
AutoSequence Editor of ES601pc, you can create any kind of General Inspection you wish, and
which may include pass/fail tests as well as entry of measurement data or remote control operation
of an external test instrument or analyzer. However, a General Inspection cannot include an
ES601
Plus
Safety AutoSequence or an ECG AutoSequence; to include safety and/or ECG tests
within the scope of another automated test requires you to set up a full Test Procedure.
Automated Testing/Chapter 6
#
Page 116
Содержание ES601 Plus
Страница 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Страница 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Страница 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Страница 59: ...ES601Plus OPERATING MANUAL Fig ure 13 Connecting to a line powered DUT Connecting To ES601 Chapter 4 Page 40 ...
Страница 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Страница 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Страница 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Страница 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Страница 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Страница 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Страница 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Страница 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Страница 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...