ES601
Plus
OPERATING MANUAL
...........................................OK
28 SET PULSE AMPL=100%
...........................................OK
29 SET ARTIFACT=OFF
...........................................OK
30 CHECK SPO2=96-98%; RATE=59-61BPM
SPO2=96% RATE=60BPM - OK...................PASS
31 CHECK SPO2=88-92%; RATE=59-61BPM
SPO2=89% RATE=60BPM - OK...................PASS
32 CHECK SPO2=78-82%; RATE=59-61BPM
SPO2=79% RATE=60BPM - OK...................PASS
33 CHECK SPO2=68-72%; RATE=59-61BPM
SPO2=68% RATE=60BPM - OK...................PASS
34 SET SPO2=93%
...........................................OK
35 CHECK SPO2=92-94%; RATE=119-121BPM
SPO2=92% RATE=119BPM - OK..................PASS
36 CHECK SPO2=92-94%; RATE=178-182BPM
SPO2=93% RATE=179BPM - OK..................PASS
37 SET HEART RATE=60BPM
...........................................OK
38 CHECK SPO2=92-94% (LOW PERFUSION)
SPO2=92%, WEAK PLETH SIGNAL................PASS
39 CHECK LOW SIGNAL ALARM
...........................................PASS
40 SET PULSE AMPL=100%
...........................................OK
41 END OF SEQUENCE
...........................................OK
42 MEASURE & RECORD DEFIB OUTPUT AT 50 J
49.8 JOULES................................PASS
43 MEASURE & RECORD DEFIB OUTPUT AT 100 J
98.4 JOULES................................PASS
44 MEASURE & RECORD DEFIB OUTPUT AT 200 J
198.1 JOULES...............................PASS
45 MEASURE & RECORD DEFIB OUTPUT AT 300 J
297.3 JOULES...............................PASS
46 MEASURE & RECORD DEFIB OUTPUT AT 360 J
352.8 JOULES...............................PASS
47 TEST MAX ENERGY WITH DEFIB ON BAT. POWER
350.2 JOULES - OK..........................PASS
48 TEST CHARGE TIME WITH DEFIB ON BAT. POWER
12 SEC - OK................................PASS
49 TEST DEFIB "SYNC MODE" OPERATION
...........................................PASS
50 CHECK PACER, DEMAND MODE OPERATION
...........................................PASS
51 MEASURE & RECORD PULSE AMPL & FREQUENCY
AMP=100MA PRR=80PPM - OK...................PASS
52 MEASURE & RECORD PULSE WIDTH
20.2 MSEC..................................PASS
53 SAF01
EXECUTE SAFETY AUTOSEQUENCE................OK
54 AFFIX "BATTERY REPLACED" LABEL IF NEEDED
...........................................OK
55 AFFIX "SERVICED" LABEL
...........................................OK
56 NOTE ANY INVENTORY OR PM CHANGES IN CEMS
...........................................OK
Appendix B
#
Page 202
Содержание ES601 Plus
Страница 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Страница 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Страница 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Страница 59: ...ES601Plus OPERATING MANUAL Fig ure 13 Connecting to a line powered DUT Connecting To ES601 Chapter 4 Page 40 ...
Страница 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Страница 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Страница 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Страница 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Страница 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Страница 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Страница 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Страница 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Страница 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...