ES601
Plus
OPERATING MANUAL
NUMBER OF LEADS:
Determines the number of leads in the group being displayed. Leads
are incorporated into the group consecutively, starting at jack AP1 on
the front panel. The simplest permitted setup is one applied part (AP1)
in one enabled group (Group 1).
BACK:
Returns to the APPLIED PARTS menu.
RESET ALL:
Resets all Group settings, leaving Group 1 enabled as Type BF with
one applied part (AP1).
If the DUT is a defibrillator and has been specified by the manufacturer as meeting the requirements
of the IEC60601-2-4 standard ("
Particular requirements for the safety of cardiac defibrillators
"), group
settings can be programmed via the GROUPS editor menu to enable special leakage tests which are
required by that standard. For testing a defibrillator per IEC60601-2-4:
• Assign ECG leads to one group, and set the GROUP TYPE to "CF ISOLATED"
• Assign the defibrillator paddles to a second group, and set NUMBER OF LEADS equal to
two. Set the GROUP TYPE to "BF ISOLATED" or "CF ISOLATED" according to
manufacturer specifications. Finally, enter "PADDLES", "PIASTRE" or "PALETTE" for
the NAME of the group.
The above special settings will enable ES601
Plus
to perform leakage testing of the paddles per
IEC60601-2-4, with modified acceptance limits according to that standard.
After configuring the leakage test conditions and, if necessary, specifying the DUT applied parts, the
next step displays the Sequence Controls for manual leakage tests, as recalled from System Settings
Menu #2 (see Section 3.3). You may change the displayed settings to suit the upcoming test. After
the test is completed, the Sequence Controls will revert to their System Setting defaults.
Manual Testing/Chapter 5
#
Page 70
Содержание ES601 Plus
Страница 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Страница 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Страница 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Страница 59: ...ES601Plus OPERATING MANUAL Fig ure 13 Connecting to a line powered DUT Connecting To ES601 Chapter 4 Page 40 ...
Страница 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Страница 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Страница 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Страница 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Страница 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Страница 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Страница 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Страница 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Страница 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...