96
Setup
AccuVision
®
AV6010 Long-Range Camera System Manual
Accu-Sort Systems, Inc.
05-AV6010_Setup_R10.doc
RangeFinder Wizard
This wizard is used to learn the RangeFinder’s position over the conveyor,
including the usable conveyor (region of interest) and height above the
conveyor surface.
Prerequisites:
Calibration box
Tape measure
Confirm the RangeFinder is centered over the scanning area
Confirm the RangeFinder is level
and that excess codewords are
blocked if the application is using a narrow belt (under 40-inches)
Laptop with Install Wizard connected to camera system
To auto-calibrate the RangeFinder:
1. Click the
RangeFinder Wizard
button.
2. Set the
Mount Selector
and click
Next
. (Normally left as
Standard
.)
3. Confirm the RangeFinder scan line is perpendicular to the conveyor’s
direction of travel using a calibration box
.
4. Learn the
Region of Interest
.
a. Measure the conveyor width used by the application.
b. Measure the usable conveyor from:
i. Left edge to the left-side RangeFinder spot.
ii. Right edge to the right-side RangeFinder spot.
c. Enter the three measurements in the wizard and click
Next
.
5. Confirm
Height Above Conveyor Surface
using a calibration box
.
6. Use the
Summary
to review the position information that will be sent to
the camera system.
7. Click
Finish
to save the RangeFinder position information to the camera
system.
8. Continue with the
Camera
wizard.
NOTE:
It is recommended that these numbers not be changed.