7.2
Hot Lips Tube Sealer
General troubleshooting
Problem
Possible cause
Corrective action
The LCD display is not
lit.
•
The instrument is not plugged
in.
•
The power switch is in the "
O"
position.
•
Instrument fuse has blown.
•
Check that the instrument is
plugged in.
•
Check that the power switch is in
the "
I" position.
•
If any instrument fuses has blown,
refer to
on how to replace fuses.
•
If problem persists, contact
Service.
The seal is not satis-
factory
•
The tubing program is not set
correctly.
•
Incorrect tubing is used.
•
The tubing is not centered in
the jaws.
•
The motor needs adjustment.
•
Check that the tubing program is
set correctly, by clicking on the
MENU button and co-relating the
tube inserted to the selected
tubing.
•
Check that you are using the
correct tubing.
•
Check that tubing is centered in the
jaws.
•
Perform mechanical assembly cali-
bration according to the instruc-
tions in
.
LCD error message:
MOTOR OVERLOAD
Mechanism was obstructed from
complete movement.
Remove tubing and any other object
from the jaws and restart the instru-
ment.
LCD error message:
HEATER OVERTEMP
The heater temperature was too
high due to internal malfunction.
Contact Service.
LCD error message:
SENSOR FAILURE
The temperature sensor has
failed, or is giving invalid read-
ings.
Contact Service.
LCD error message:
HEATER FAILURE
The heater did not reach set
temperature within time-out
period.
Contact Service.
7 Troubleshooting
7.2 Hot Lips Tube Sealer
Sterile Tube Fuser - Dry and Hot Lips Tube Sealer Operating Instructions 28964139 AG
91