Step
Action
4
Clean the spillage area on the instrument surface with a proper cleaning
agent and cotton swab.
WARNING
Do not attempt to touch the instrument in the power ON condition,
if large amount of liquid spillage (approx. > 50 mL) is found inside
the instrument. Switch off the instrument, disconnect the power
cord and contact Service.
Lid
Clean the polycarbonate lid with 40% ethanol or 40% isopropanol.
Danger zone
The danger zone under the cover see
Section 2.4 Danger zone, on page 27
cleaned periodically, using a small handheld vacuum cleaner. Any spillage can be
cleaned with a lint free cloth moistened with water or up to 20% ethanol.
Infrared sensor window
Clean the infrared sensor window (indicated by an arrow below) every 2 months. Use a
lens cleaning kit to clean the sensor window.
6 Maintenance
6.2 Cleaning the instrument
6.2.1 Cleaning of Sterile Tube Fuser - Dry
66
Sterile Tube Fuser - Dry and Hot Lips Tube Sealer Operating Instructions 28964139 AG