4.1
Site requirements
Introduction
This section describes the requirements for space and supplies in the building where
the Sterile Tube Fuser - Dry and Hot Lips Tube Sealer instruments will be installed.
Conditions
Sterile Tube Fuser - Dry
Parameter
Requirements
Allowed location
Indoor use only
Electrical power
100-120/220-240 V~, 50/60 Hz
Transient overvoltages
Overvoltage category II
Operating temperature
0°C to 30°C
Humidity
20% to 80%, non-condensing
Pollution degree
2
Altitude
Maximum 2000 m
Hot Lips Tube Sealer
Parameter
Requirements
Allowed location
Indoor use only
Electrical power
100-120/220-240 V~, 50/60 Hz
Transient overvoltages
Overvoltage category II
Operating temperature
2°C to 32°C
Humidity
2% to 80%, non-condensing
Pollution degree
2
Altitude
Maximum 2000 m
Space requirements
The following requirements apply to both Sterile Tube Fuser - Dry and Hot Lips Tube
Sealer.
4 Installation
4.1 Site requirements