![cytiva Sterile Tube Fuser Скачать руководство пользователя страница 71](http://html1.mh-extra.com/html/cytiva/sterile-tube-fuser/sterile-tube-fuser_operating-instructions-manual_2706957071.webp)
6.4.1
Tube Restrictor replacement
Introduction
The Tube Restrictors are wearable components that require periodic replacement to
ensure good quality sealing of thermoplastic tubing.
Depending on tube size and tubing formulation, approximately 40 to 60 seals can be
created before the Tube Restrictors need to be replaced. It is recommended to replace
the Tube Restrictors as a part of the routine maintenance of Hot Lips Tube Sealer.
For more information about inspecting the quality of the seal, see
Tools required
•
Torx™ wrenches T8 and T10
•
Needle nosed pliers
Tube Restrictor assembly illustration
4
1
2
3
5 6
Part
Description
1
Red flag indicator
2
Metal steering peg
3
Hex pan head screw(1)
4
Hex pan head screw(2)
6 Maintenance
6.4 Maintenance of Hot Lips Tube Sealer
6.4.1 Tube Restrictor replacement
Sterile Tube Fuser - Dry and Hot Lips Tube Sealer Operating Instructions 28964139 AG
71