Error No.
Message
Possible cause
Corrective action
ERROR 10
COVER OPEN
Cover opened during opera-
tion.
A safety interlock stops the weld
cycle if the cover is forced open
or cover lock fails. Contact
Service to repair cover lock.
ERROR 11
PUSH MOTOR
FAIL
The X-axis motor
compresses the tubes
together during the weld.
This error indicates that this
motor failed to move
correctly.
•
Switch off the power to the
instrument.
•
Switch on the power to the
instrument and follow the
display messages.
•
If the motor fails to move,
contact Service.
ERROR 12
RESET FAIL
The instrument failed to
reset to the start configura-
tions.
Check that all tubes are removed,
and that nothing is obstructing
the movement of the motors.
ERROR 13
REMOVE TUBES
The instrument was
powered up with tubes still
inserted in the holders.
Remove all tubes and restart the
instrument.
N/A
ERFF
•
Time out error.
•
The communication
cable may be discon-
nected, or the communi-
cation error has
occurred due to noise
and so on.
•
Check if the cable is
connected properly.
•
Switch off the power to the
instrument.
•
Wait for 2 to 3 minutes.
•
Switch on the power to the
instrument.
N/A
ERFE
Response error from the PC
during communication with
the instrument.
•
Verify that the expected data
is received from the PC (GUI
application).
•
Check that the communica-
tion cable is connected prop-
erly.
•
Restart the PC application
(GUI).
7 Troubleshooting
7.1 Sterile Tube Fuser - Dry
90
Sterile Tube Fuser - Dry and Hot Lips Tube Sealer Operating Instructions 28964139 AG