![cytiva Sterile Tube Fuser Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/cytiva/sterile-tube-fuser/sterile-tube-fuser_operating-instructions-manual_2706957051.webp)
The Tubing holders are coded so that the instrument automatically reads the Tubing
holder type and selects the appropriate program.
Step
Action
1
Open the cover.
2
For each Tubing holder:
•
Push the Locking tab towards the rear of the instrument to unlock the
Tubing holder (arrows).
•
Lift out the Tubing holder.
3
Select desired Tubing holder pair (the pair must match).
4
Place each Tubing holder on its platform.
Note:
Ensure that the alignment pins under the Tubing holder are placed into holes
in the platform.
5
Pull the locking tabs towards the front of the instrument to lock the Tubing
holders.
Handling of liquid filled tubing
CAUTION
Do not attempt to connect liquid-filled tubing using Sterile Tube
Fuser - Dry, or allow liquid to spill into the instrument.
5 Operation
5.2 Sterile Tube Fuser - Dry
5.2.2 Fusing tubes
Sterile Tube Fuser - Dry and Hot Lips Tube Sealer Operating Instructions 28964139 AG
51