Step
Action
6
•
Press and hold the
GO button.
•
Release the
GO button when a beep is heard.
If the
GO button is released before the jaws have closed to the preset
sealing position, the jaws open again.
The jaws close to squeeze the tubing. The
READY light goes out. Once the
jaws are closed, the sealing begins.
The LCD display shows the progress of the cycle and the time in seconds for
each step:
Step
LCD display
Step
LCD display
1
4
2
5
3
6
When the seal cycle is complete, the instrument opens to the release posi-
tion and the sealing jaws are cool enough to touch.
7
When the
READY light turns on, remove the tubing.
8
If another tube sealing is desired, press the
GO button and the instrument
resets. Follow the instructions (Sealing tubes) from step 3 and onwards.
Note:
Each seal cycle takes about 2 to 3 minutes depending on the tubing type and
dimension.
9
Inspect the seal carefully before use, see
Inspect seal quality on page 59
5 Operation
5.3 Hot Lips Tube Sealer
58
Sterile Tube Fuser - Dry and Hot Lips Tube Sealer Operating Instructions 28964139 AG