Start instrument and set tubing
brand and dimension
Step
Action
1
Switch on the power to Sterile Tube Fuser - Dry by pressing the power switch
to the "
I" position.
Result:
Close the lid and lock by pressing cover lock button in order to proceed
further.
On successful initialization, the display shows the
INSERT BLADE screen.
1
5
4
3
2
The display shows current date and time (1) and instrument ID number (4).
2
Check that desired tubing dimension (2) and tubing brand (3) are shown.
3
Optional, change tubing brand (3):
•
Press the
CHANGE button (5).
•
Press the desired tubing type on the new display. If no tubing types
appear on the first screen, press the
NEXT button to open the second
screen.
Note:
If you have already inserted the blade and wish to change the tubing type,
simply remove the blade to get the
INSERT BLADE screen.
Note:
If the tubing has been inserted you must remove the tubing, restart the
instrument, and then remove the blade to get the
INSERT BLADE screen.
4
Optional, change tubing dimension:
Use a Tubing holder with the desired dimension. The tubing dimension is
then automatically entered in the instrument.
To change a Tubing holder, see
Changing Tubing holder on page 50
.
5 Operation
5.2 Sterile Tube Fuser - Dry
5.2.2 Fusing tubes