Step
Action
3
Remove the used cutting blade with the Installation/Removal Tool.
Secure the blade within the tool by compressing with the thumb as shown in
the image below.
Note:
For optimal operation, do not reuse cutting blades.
4
Open the red Tubing holder, then the blue Tubing holder.
5
Remove the connected tubing and discard the cut ends.
Note:
Inspect the weld carefully before use, see
.
6
Close the cover and press the cover lock button to reset the instrument.
7
Start instrument and set tubing brand and dimension on
to make additional tubing welds.
Each weld cycle takes about 2 minutes.
8
Switch off the power to Sterile Tube Fuser - Dry by pressing the power switch
to the "
O" position.
Changing Tubing holder
The Tubing holders can be quickly changed to connect different diameter tubing. The
Tubing holders must be changed as a pair (left and right Tubing holders must be the
same size).
5 Operation
5.2 Sterile Tube Fuser - Dry
5.2.2 Fusing tubes
50
Sterile Tube Fuser - Dry and Hot Lips Tube Sealer Operating Instructions 28964139 AG