Step
Action
5
If...
then...
only the PASS end of
the Functional
verification gauge
passes through the
jaws without any
difficulty
press the
Cursor up button.
Result: The display shows
CAL PASSED AND
RECYCLE POWER. Restart the instrument.
any part of the FAIL
end of the
Functional
verification gauge
passes through the
jaws
press the
Cursor down button.
Result: The display shows
CALIBRATION FAIL.
Calibrate to make sure that jaws are within the
specified tolerance.
Calibration
Follow the steps below to calibrate Hot Lips Tube Sealer.
Step
Action
1
Switch the power switch to the "
O" position.
6 Maintenance
6.4 Maintenance of Hot Lips Tube Sealer
6.4.2 Functional validation and calibration
Sterile Tube Fuser - Dry and Hot Lips Tube Sealer Operating Instructions 28964139 AG