Operation Manual
NORM-O-TEMP
®
, Model 111Z
Page 43 of 50
B.
Remove the three sided enclosure panel.
C.
Locate the water filter assembly.
D.
Firmly grasp the notched rim of the cap of the assembly and turn it counterclockwise. The
cap may contain a small amount of water.
E.
Remove the wire mesh.
F.
Clean the wire mesh and the plastic cap. Be careful not to lose the black O-ring in the rim
of the plastic cap.
G.
Replace the wire mesh in the plastic cap and position the wire mesh, O-ring and plastic
cap.
H.
Screw the cap clockwise onto the fitting until it is secure.
I.
Replace the three sided enclosure panel.
J.
Replenish the water reservoir as described in Section 5-1.5.
5-3.
Maintenance of the NORM-O-TEMP® Exterior & Vents – Cleaning Instructions
The NORM-O-TEMP
®
unit is constructed of powder coated steel with a plastic top that should be
cleaned quarterly with water and mild detergent solution. For cleaning and disinfecting, always
use conventional hospital-approved topical equipment cleaners and disinfectants that do not
contain alcohol. Avoid alcohol and other strong, undiluted disinfectants. These may cause
staining of the device’s outer skin.
Thoroughly wipe down device with a damp cloth to remove any residue from cleaning solutions.
Ensure that all inaccessible cracks and crevices on the NORM-O-TEMP
®
unit are reached during
cleaning.
Wipe the vents on the front and rear of the unit. Ensure all debris is removed and there is no
blockage of any vent on the unit. Vents must be free of obstructions.
In addition, ensure that all accessories associated with the NORM-O-TEMP
®
system are cleaned
quarterly as well.
WARNING
•
The repair, calibration, and servicing of the NORM-O-TEMP
®
should be performed by qualified
Medical Equipment Service Technicians, Certified Biomedical Electronics Technicians, or
Certified Clinical Engineers familiar with good repair practices for servicing medical devices,
and in accordance with instructions contained in this manual.
Improper repair can result in
damage to the NORM-O-TEMP
®
system and patient injury.
•
Always
unplug the unit
before accessing internal components during service. Failure to
unplug the unit could result in
electric shock.
Содержание Norm-o-Temp 111Z
Страница 13: ...Operation Manual NORM O TEMP Model 111Z Page 13 of 50 Figure 1 NORM O TEMP unit Front View...
Страница 15: ...Operation Manual NORM O TEMP Model 111Z Page 15 of 50 Figure 2 NORM O TEMP unit Left Side View...
Страница 17: ...Operation Manual NORM O TEMP Model 111Z Page 17 of 50 Figure 3 NORM O TEMP unit Rear View...
Страница 19: ...Operation Manual NORM O TEMP Model 111Z Page 19 of 50 Figure 4 NORM O TEMP unit Right Side View...
Страница 21: ...Operation Manual NORM O TEMP Model 111Z Page 21 of 50 Figure 5 NORM O TEMP unit Model 111Z Top View...
Страница 49: ...NORM O TEMP Model 111Z Page 49 of 50 THIS PAGE INTENTIONALLY LEFT BLANK...