background image

Operation Manual                                

   NORM-O-TEMP

®

, Model 111Z  

 
 

Page 18 of 50 

 

1-3.4.

 

External Features and Descriptions – Right Side View 

The external features in Figure 4 of the NORM-O-TEMP

®

 unit are described as follows: 

 
A.

 

The three sided enclosure panel, secured with eight screws, provides access to the 
interior. 

 
B.

 

The recessed handle is one of two provided to assist in lifting the unit. 

 

C.

 

Two screws on the right and left side of the unit secure the top to the base. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание Norm-o-Temp 111Z

Страница 1: ...Operation Manual Model 111Z Hyperthermia System Cincinnati Sub Zero Products LLC 12011 Mosteller Road Cincinnati Ohio 45241 U S A www cszmedical com...

Страница 2: ...50 NORM O TEMP MAXI THERM MAXI THERM LITE Gelli Roll and PLASTI PAD are registered trademarks of Cincinnati Sub Zero Products LLC Cincinnati Ohio USA Copyright 2016 Cincinnati Sub Zero Products LLC Al...

Страница 3: ...Temperature Decreasing Temperature Disconnect Power Before Servicing Water Flow Indicator Outlet Inlet Silence Alarm Caution Read Operation Instructions and Manual Before Operating Temp Set Point Ind...

Страница 4: ...rticularly the skin to a point where they are injured Depending on the extent and severity of a burn very serious and even fatal complications may arise Do not use the NORM O TEMP System distal to art...

Страница 5: ...P SYSTEM TEST ROUTINE in the Operation Technical Manual must always be performed Improper repair and inadequate maintenance can result in damage to the NORM O TEMP system and patient injury Remove the...

Страница 6: ...er Electrostatic Discharge ESD practices should be followed during replacement of any electronic board To Discharge the NORM O TEMP system ensure the unit is unplugged and use any metal tool such as a...

Страница 7: ...m 26 3 0 Introduction 26 3 1 Unpacking the Shipment 26 3 2 First Time Set Up System Test Routine 26 3 2 1 Inspecting and Arranging the Equipment 26 3 2 2 Completing a System Test Routine 28 3 3 Unit a...

Страница 8: ...Figure 2 NORM O TEMP unit Left Side View 15 Figure 3 NORM O TEMP unit Rear View 17 Figure 4 NORM O TEMP unit Right Side View 19 Figure 5 NORM O TEMP unit Model 111Z Top View 21 Figure 6 NORM O TEMP Mo...

Страница 9: ...vailable at the time of purchase or during the warranty period To return defective parts or units first obtain a Returned Materials Authorization RMA number from our Medical Technical Service departme...

Страница 10: ...NOTE You have five seconds to depress one of the arrows or the temperature readout will return to the actual reservoir water temperature 6 Depress the UP or DOWN arrow to raise or lower the set point...

Страница 11: ...vice personnel who are responsible for maintaining the equipment All personnel who operate or service the unit should be familiar with all parts of this manual Physical and functional characteristics...

Страница 12: ...M O TEMP System See Section 2 for specifications and certifications of the NORM O TEMP system 1 3 1 External Features and Descriptions Front View The external features in Figure 1 of the NORM O TEMP u...

Страница 13: ...Operation Manual NORM O TEMP Model 111Z Page 13 of 50 Figure 1 NORM O TEMP unit Front View...

Страница 14: ...ricted The change in water flow decreases the speed of the paddle wheel The water flow indicator only spins when a blanket pad or by pass hose is connected to the unit It will not spin when water is c...

Страница 15: ...Operation Manual NORM O TEMP Model 111Z Page 15 of 50 Figure 2 NORM O TEMP unit Left Side View...

Страница 16: ...ir vents provide air circulation for the internal components C The nylon straps are used to secure and store the coiled power cord and or the connecting hose when not in use D The serial number printe...

Страница 17: ...Operation Manual NORM O TEMP Model 111Z Page 17 of 50 Figure 3 NORM O TEMP unit Rear View...

Страница 18: ...nal features in Figure 4 of the NORM O TEMP unit are described as follows A The three sided enclosure panel secured with eight screws provides access to the interior B The recessed handle is one of tw...

Страница 19: ...Operation Manual NORM O TEMP Model 111Z Page 19 of 50 Figure 4 NORM O TEMP unit Right Side View...

Страница 20: ...temperature reaches 46 0 C 0 6 C 115 F 1 08 F D When the LO WATER LED is lit it indicates that the unit needs more distilled water to operate The LED will be accompanied by an audible alarm E The SIL...

Страница 21: ...Operation Manual NORM O TEMP Model 111Z Page 21 of 50 Figure 5 NORM O TEMP unit Model 111Z Top View...

Страница 22: ...ters capacity dual compartment reservoir is composed of the circulating reservoir situated under and connected to the replenishing reservoir When the operator fills the reservoir with distilled water...

Страница 23: ...t from damage caused by extreme high temperatures Safety Controls for Protection from High Temperature If the circulating water reaches 42 0 C 0 6 C 107 6 F 1 08 F the microprocessor shuts off heater...

Страница 24: ...al Greater than 1 Over Set point Audible and Visual Control System Microprocessor based temperature control system and Alarm indications Option to select temperature settings as either C or F Controll...

Страница 25: ...itional Warranty Available Safety System Maximum High Control Setting 42 C 107 6 F Primary Temperature Limit 43 5 C 0 6 C 110 0 F 1 08 F Secondary Temperature Limit 44 5 C 0 6 C 112 0 F 1 08 F Indepen...

Страница 26: ...found notify the transportation company immediately and file a claim The transportation company is responsible for the shipment after it leaves the factory If problems other than shipping damage are...

Страница 27: ...RM O TEMP unit Hand tighten knurled knobs until snug G Slide the lid of the water fill opening and add 340 cc of propylene glycol then gradually pour approximately 6 quarts 5 7 liters of distilled wat...

Страница 28: ...level the LO WATER LED is activated and the alarm sounds The unit shuts down and the operator cannot proceed until this is corrected by filling the reservoir A Make sure that the power switch is in th...

Страница 29: ...sent a risk of infection to the patient because the water is not sterile If a leak is found the blanket pad should not be used I Check the couplings at the unit and at the blanket pad for positive con...

Страница 30: ...the prescribed temperature range B Changes in skin color edema inflammation or indications of pressure especially over bony prominences should be noted and treated as ordered Avoid prolonged tissue pr...

Страница 31: ...ical high temperature safety will activate to shut off the power to the pump and heater In addition the HI TEMP LED and Power Fail LED illuminate and the audible alarm sounds 2 If the independent mech...

Страница 32: ...fill opening and check if the water is visibly touching the strainer If needed carefully add distilled water Do not use de ionized water Do not overfill If the water falls below a preset level the al...

Страница 33: ...nt preparation and bedside care are further described in Section 3 4 NOTE If at anytime power is removed from the unit without actuating the power switch I O switch a power fail alarm will activate an...

Страница 34: ...ater to drain back into the unit B Disconnect the power cord from the power source loosely coil it and attach it to the back panel using the nylon straps C Remove the blanket pad s D Loosely coil the...

Страница 35: ...cautions necessary to complete the requirements Maintenance repair and or service of NORM O TEMP by qualified maintenance technical personnel will not void the warranty of the unit The following maint...

Страница 36: ...ater See Section 5 1 5 10 Check flow rate a Flow rate Minimum of 0 5 GPM check both sets of fittings b Pressure Maximum of 10 PSI check both sets of fittings 11 Condition of blanket pad s hoses and co...

Страница 37: ...TEMP 111Z Heating 38 C 5 Minutes 5 1 1 Internal Cleaning using Bleach Dry Storage Procedure For circulating water units utilizing household bleach The following procedure cleans the fluid circuit in t...

Страница 38: ...t step M K If unit is to be returned to service once no bleach is detected add the appropriate amount of U S P Grade Propylene Glycol to the water reservoir per the following chart UNIT U S P Grade Pr...

Страница 39: ...reservoir with distilled water E Turn the unit on and circulate per the Chemical Cleaning Circulation Chart above F Drain the unit as instructed in Step A G Rinse the unit with Distilled Water three 3...

Страница 40: ...amount of water per the above chart E Turn the unit on and circulate per the Chemical Cleaning Circulation Chart above F Drain the unit as instructed in Step A G Rinse the unit three 3 times as descr...

Страница 41: ...r switch to the I position The water is pumped into the container until there is approximately 1 quarts 1 4 liters of water remaining to be drained The LO WATER indicator lights up the status display...

Страница 42: ...ped fitting that intersects the hose from the water flow indicator to the reservoir A stainless steel wire mesh screen is located inside the clear plastic cap of the water filter assembly Quarterly or...

Страница 43: ...rs and disinfectants that do not contain alcohol Avoid alcohol and other strong undiluted disinfectants These may cause staining of the device s outer skin Thoroughly wipe down device with a damp clot...

Страница 44: ...zation or autoclaving for cleaning and disinfecting the blankets pads To drain the water from the reusable blankets pads simply shut off power to the unit and allow the water to drain from the blanket...

Страница 45: ...3 Infant Size 12 5 x 18 872 Pediatric 25 x 19 871 Pediatric 25 x 17 25 870 Localized therapy 4 x 25 300 Staff vest PLASTIPAD Reusable Molded Plastic Blanket pad s Polyurethane 196 Adult Size 24 x 60 1...

Страница 46: ...l Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors a...

Страница 47: ...ration distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference...

Страница 48: ...Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0...

Страница 49: ...NORM O TEMP Model 111Z Page 49 of 50 THIS PAGE INTENTIONALLY LEFT BLANK...

Страница 50: ...Cincinnati Sub Zero Products LLC 12011 Mosteller Road Cincinnati OH 45241 Telephone 1 800 989 7373 or 513 772 8810 Fax 513 772 9119 E Mail csz genthermcsz com...