   NORM-O-TEMP, Model 111Z  

 

 

Page 4 of 50 

Cincinnati Sub-Zero Products, LLC, reserves the right to make changes to the device, which may not be 
reflected in this manual.

 

WARNING 

 

A physician's order is required for the use of the device and setting the temperature of the 
blanket/pad.  At least every 20 minutes, or as directed by physician, check patient's 
temperature and skin condition of areas in contact with blanket/pad; also, check blanket/pad 
water temperature.  Pediatric patients, temperature-sensitive patients with vascular disease, 
surgical patients, diabetic and Reynaud’s disease patients should be checked more frequently.  If 
patient’s temperature does not reach desired set point or differs drastically from recommended 
set point, notify physician.  

Notify the physician promptly of any change in patient status in 

order to avoid serious injury or death.  

 

1.

 

Pediatrics – Infants’ and children’s body temperatures are often more responsive to 
surface heating and cooling than adults.  Due to their size, the effect of heating or 
cooling a child is likely more pronounced because of their higher ratio of skin contact 
area to body mass. 

2.

 

Temperature Sensitive Patients – Patients with impaired peripheral blood circulation 
and patients who are incapacitated may be more sensitive to temperature changes 
than patients with normal circulation. 

3.

 

Surgical Patients – Patients with poor circulation due to inadequate heart function, loss 
of blood, or impaired peripheral blood circulation may be more sensitive to 
temperature changes. 

 

The method of temperature control provided by all hyperthermia units presents the danger of 
heating body tissues, particularly the skin, to a point where they are injured.  Depending on the 
extent and severity of a burn, 

very serious and even fatal complications may arise.  

 

 

Do not use the NORM-O-TEMP

®

 System distal to arterial cross clamping.  

Thermal injury may 

result.

 

 

Prevent excessive and/or prolonged tissue pressure and shearing forces, especially over bony 
prominences, to prevent 

skin

 

damage that may result.

 

 

Do not place additional heat sources between the patient and blanket/pad.  

Skin damage may 

result.

 

 

The area between the patient and the blanket/pad should be kept dry to avoid injury to 
patient.  

Prep solutions have been reported to injure the skin when allowed to remain between 

patients and a water-circulating heating blanket/pad during prolonged procedures.   

 

Proper sanitation procedures must be practiced and hygienic safety must be maintained, 
to prevent contamination.  

Contamination can affect patient’s health, i.e.  skin irritation/rash 

may result.

 

 

Do not use the NORM-O-TEMP

®

 system in the presence of flammable anesthetics.  

Risk of 

explosion can result.

 

 

The warming of transdermal medications (patches) can 

increase drug delivery, resulting in 

possible injury to the patient.

 

 

Power interruption will cause the NORM-O-TEMP

®

 unit to revert to Preset Set point Temperature 

resulting  in  possible  inaccurate  therapy  to  the  patient.    Follow  instructions  for  First  Time  Set-
Up/System  Test  Routine  located  in  the  Operation/Technical  Manual  to  resume  operation.  

Failure to resume therapy could result in serious injury or death. 

 

Do not by-pass ground lug.  

 Electrical Hazards may result.

 

 

Any time water is found leaking into or around the unit, connecting hose, and/or blanket/pad, 
turn  the  unit  off,  disconnect  the  power  cord  from  its  power  source,  and  correct  the  problem 
before  proceeding.   

Water  leaks  could  lead  to  electric  shock.    Water  leaks  could  also 

present a slip hazard.  

 

Содержание Norm-o-Temp 111Z

Страница 1: ...Operation Manual Model 111Z Hyperthermia System Cincinnati Sub Zero Products LLC 12011 Mosteller Road Cincinnati Ohio 45241 U S A www cszmedical com...

Страница 2: ...50 NORM O TEMP MAXI THERM MAXI THERM LITE Gelli Roll and PLASTI PAD are registered trademarks of Cincinnati Sub Zero Products LLC Cincinnati Ohio USA Copyright 2016 Cincinnati Sub Zero Products LLC Al...

Страница 3: ...Temperature Decreasing Temperature Disconnect Power Before Servicing Water Flow Indicator Outlet Inlet Silence Alarm Caution Read Operation Instructions and Manual Before Operating Temp Set Point Ind...

Страница 4: ...rticularly the skin to a point where they are injured Depending on the extent and severity of a burn very serious and even fatal complications may arise Do not use the NORM O TEMP System distal to art...

Страница 5: ...P SYSTEM TEST ROUTINE in the Operation Technical Manual must always be performed Improper repair and inadequate maintenance can result in damage to the NORM O TEMP system and patient injury Remove the...

Страница 6: ...er Electrostatic Discharge ESD practices should be followed during replacement of any electronic board To Discharge the NORM O TEMP system ensure the unit is unplugged and use any metal tool such as a...

Страница 7: ...m 26 3 0 Introduction 26 3 1 Unpacking the Shipment 26 3 2 First Time Set Up System Test Routine 26 3 2 1 Inspecting and Arranging the Equipment 26 3 2 2 Completing a System Test Routine 28 3 3 Unit a...

Страница 8: ...Figure 2 NORM O TEMP unit Left Side View 15 Figure 3 NORM O TEMP unit Rear View 17 Figure 4 NORM O TEMP unit Right Side View 19 Figure 5 NORM O TEMP unit Model 111Z Top View 21 Figure 6 NORM O TEMP Mo...

Страница 9: ...vailable at the time of purchase or during the warranty period To return defective parts or units first obtain a Returned Materials Authorization RMA number from our Medical Technical Service departme...

Страница 10: ...NOTE You have five seconds to depress one of the arrows or the temperature readout will return to the actual reservoir water temperature 6 Depress the UP or DOWN arrow to raise or lower the set point...

Страница 11: ...vice personnel who are responsible for maintaining the equipment All personnel who operate or service the unit should be familiar with all parts of this manual Physical and functional characteristics...

Страница 12: ...M O TEMP System See Section 2 for specifications and certifications of the NORM O TEMP system 1 3 1 External Features and Descriptions Front View The external features in Figure 1 of the NORM O TEMP u...

Страница 13: ...Operation Manual NORM O TEMP Model 111Z Page 13 of 50 Figure 1 NORM O TEMP unit Front View...

Страница 14: ...ricted The change in water flow decreases the speed of the paddle wheel The water flow indicator only spins when a blanket pad or by pass hose is connected to the unit It will not spin when water is c...

Страница 15: ...Operation Manual NORM O TEMP Model 111Z Page 15 of 50 Figure 2 NORM O TEMP unit Left Side View...

Страница 16: ...ir vents provide air circulation for the internal components C The nylon straps are used to secure and store the coiled power cord and or the connecting hose when not in use D The serial number printe...

Страница 17: ...Operation Manual NORM O TEMP Model 111Z Page 17 of 50 Figure 3 NORM O TEMP unit Rear View...

Страница 18: ...nal features in Figure 4 of the NORM O TEMP unit are described as follows A The three sided enclosure panel secured with eight screws provides access to the interior B The recessed handle is one of tw...

Страница 19: ...Operation Manual NORM O TEMP Model 111Z Page 19 of 50 Figure 4 NORM O TEMP unit Right Side View...

Страница 20: ...temperature reaches 46 0 C 0 6 C 115 F 1 08 F D When the LO WATER LED is lit it indicates that the unit needs more distilled water to operate The LED will be accompanied by an audible alarm E The SIL...

Страница 21: ...Operation Manual NORM O TEMP Model 111Z Page 21 of 50 Figure 5 NORM O TEMP unit Model 111Z Top View...

Страница 22: ...ters capacity dual compartment reservoir is composed of the circulating reservoir situated under and connected to the replenishing reservoir When the operator fills the reservoir with distilled water...

Страница 23: ...t from damage caused by extreme high temperatures Safety Controls for Protection from High Temperature If the circulating water reaches 42 0 C 0 6 C 107 6 F 1 08 F the microprocessor shuts off heater...

Страница 24: ...al Greater than 1 Over Set point Audible and Visual Control System Microprocessor based temperature control system and Alarm indications Option to select temperature settings as either C or F Controll...

Страница 25: ...itional Warranty Available Safety System Maximum High Control Setting 42 C 107 6 F Primary Temperature Limit 43 5 C 0 6 C 110 0 F 1 08 F Secondary Temperature Limit 44 5 C 0 6 C 112 0 F 1 08 F Indepen...

Страница 26: ...found notify the transportation company immediately and file a claim The transportation company is responsible for the shipment after it leaves the factory If problems other than shipping damage are...

Страница 27: ...RM O TEMP unit Hand tighten knurled knobs until snug G Slide the lid of the water fill opening and add 340 cc of propylene glycol then gradually pour approximately 6 quarts 5 7 liters of distilled wat...

Страница 28: ...level the LO WATER LED is activated and the alarm sounds The unit shuts down and the operator cannot proceed until this is corrected by filling the reservoir A Make sure that the power switch is in th...

Страница 29: ...sent a risk of infection to the patient because the water is not sterile If a leak is found the blanket pad should not be used I Check the couplings at the unit and at the blanket pad for positive con...

Страница 30: ...the prescribed temperature range B Changes in skin color edema inflammation or indications of pressure especially over bony prominences should be noted and treated as ordered Avoid prolonged tissue pr...

Страница 31: ...ical high temperature safety will activate to shut off the power to the pump and heater In addition the HI TEMP LED and Power Fail LED illuminate and the audible alarm sounds 2 If the independent mech...

Страница 32: ...fill opening and check if the water is visibly touching the strainer If needed carefully add distilled water Do not use de ionized water Do not overfill If the water falls below a preset level the al...

Страница 33: ...nt preparation and bedside care are further described in Section 3 4 NOTE If at anytime power is removed from the unit without actuating the power switch I O switch a power fail alarm will activate an...

Страница 34: ...ater to drain back into the unit B Disconnect the power cord from the power source loosely coil it and attach it to the back panel using the nylon straps C Remove the blanket pad s D Loosely coil the...

Страница 35: ...cautions necessary to complete the requirements Maintenance repair and or service of NORM O TEMP by qualified maintenance technical personnel will not void the warranty of the unit The following maint...

Страница 36: ...ater See Section 5 1 5 10 Check flow rate a Flow rate Minimum of 0 5 GPM check both sets of fittings b Pressure Maximum of 10 PSI check both sets of fittings 11 Condition of blanket pad s hoses and co...

Страница 37: ...TEMP 111Z Heating 38 C 5 Minutes 5 1 1 Internal Cleaning using Bleach Dry Storage Procedure For circulating water units utilizing household bleach The following procedure cleans the fluid circuit in t...

Страница 38: ...t step M K If unit is to be returned to service once no bleach is detected add the appropriate amount of U S P Grade Propylene Glycol to the water reservoir per the following chart UNIT U S P Grade Pr...

Страница 39: ...reservoir with distilled water E Turn the unit on and circulate per the Chemical Cleaning Circulation Chart above F Drain the unit as instructed in Step A G Rinse the unit with Distilled Water three 3...

Страница 40: ...amount of water per the above chart E Turn the unit on and circulate per the Chemical Cleaning Circulation Chart above F Drain the unit as instructed in Step A G Rinse the unit three 3 times as descr...

Страница 41: ...r switch to the I position The water is pumped into the container until there is approximately 1 quarts 1 4 liters of water remaining to be drained The LO WATER indicator lights up the status display...

Страница 42: ...ped fitting that intersects the hose from the water flow indicator to the reservoir A stainless steel wire mesh screen is located inside the clear plastic cap of the water filter assembly Quarterly or...

Страница 43: ...rs and disinfectants that do not contain alcohol Avoid alcohol and other strong undiluted disinfectants These may cause staining of the device s outer skin Thoroughly wipe down device with a damp clot...

Страница 44: ...zation or autoclaving for cleaning and disinfecting the blankets pads To drain the water from the reusable blankets pads simply shut off power to the unit and allow the water to drain from the blanket...

Страница 45: ...3 Infant Size 12 5 x 18 872 Pediatric 25 x 19 871 Pediatric 25 x 17 25 870 Localized therapy 4 x 25 300 Staff vest PLASTIPAD Reusable Molded Plastic Blanket pad s Polyurethane 196 Adult Size 24 x 60 1...

Страница 46: ...l Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors a...

Страница 47: ...ration distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference...

Страница 48: ...Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0...

Страница 49: ...NORM O TEMP Model 111Z Page 49 of 50 THIS PAGE INTENTIONALLY LEFT BLANK...

Страница 50: ...Cincinnati Sub Zero Products LLC 12011 Mosteller Road Cincinnati OH 45241 Telephone 1 800 989 7373 or 513 772 8810 Fax 513 772 9119 E Mail csz genthermcsz com...

Отзывы: