General Preparation of the NORM-O-TEMP
®
System
3-0.
Introduction
This section describes the procedures to prepare the NORM-O-TEMP
®
unit for general use. This
entails unpacking the shipment, arranging all the equipment for the first time, and completing a
test routine. This section also outlines the cleaning agent preparations, standard safety
precautions, and patient preparation/bedside care when using the hyper-hypothermia
blanket/pad(s). Technical support is available upon request.
3-1.
Unpacking the Shipment
WARNING
•
Remove the NORM-O-TEMP
®
from service if the outer casing or membrane control panel is
cracked or internal components are exposed.
Contact with internal components could
result in electric shock or thermal injury to the patient or operator and exposure to
sharp edges.
Open the top of the carton and remove the top foam insert. Lift the unit from the box. Locate
and remove the manual.
During the unpacking process, look carefully for signs of shipping damage. If any unacceptable
damage is found, notify the transportation company immediately and file a claim. The
transportation company is responsible for the shipment after it leaves the factory. If problems
other than shipping damage are found, notify your Cincinnati Sub-Zero representative or
manufacturer (Cincinnati Sub-Zero Products LLC) directly.
3-2.
First Time Set-Up / System Test Routine
This section describes the tasks necessary to inspect and arrange the equipment for the first
time after unpacking and describes a System Test Routine to check out the membrane control
panel.
The System Test Routine can also be used to teach operators unfamiliar with the equipment
how to use the unit.
3-2.1.
Inspecting and Arranging the Equipment
A.
Place the NORM-O-TEMP
®
unit on an uncluttered work surface that is accessible to the
correct power source. Position the unit so that the membrane control panel faces the
operator.
OPTION: The NORM-O-TEMP
®
unit can be mounted on an optional Low Profile stand
with casters or an optional I.V. Stand. For installation instructions, see insert packaged
with the stand.
B.
Visually inspect the NORM-O-TEMP
®
unit to determine that there are no missing parts,
dents, or punctures.
C.
Examine the power cord for cuts or exposed wires and the power plug for bent or
missing prongs.
Содержание Norm-o-Temp 111Z
Страница 13: ...Operation Manual NORM O TEMP Model 111Z Page 13 of 50 Figure 1 NORM O TEMP unit Front View...
Страница 15: ...Operation Manual NORM O TEMP Model 111Z Page 15 of 50 Figure 2 NORM O TEMP unit Left Side View...
Страница 17: ...Operation Manual NORM O TEMP Model 111Z Page 17 of 50 Figure 3 NORM O TEMP unit Rear View...
Страница 19: ...Operation Manual NORM O TEMP Model 111Z Page 19 of 50 Figure 4 NORM O TEMP unit Right Side View...
Страница 21: ...Operation Manual NORM O TEMP Model 111Z Page 21 of 50 Figure 5 NORM O TEMP unit Model 111Z Top View...
Страница 49: ...NORM O TEMP Model 111Z Page 49 of 50 THIS PAGE INTENTIONALLY LEFT BLANK...