Operation Manual
NORM-O-TEMP
®
, Model 111Z
Page 42 of 50
5-1.5.
Replenishing the Reservoir / Fixing a Low Water Alarm
A.
Check that the drain hose is disconnected.
B.
Slide the water fill opening lid to the left and gradually pour approximately 1½ gallons
(5.7 liters) of distilled water with approximately 340 cc of propylene glycol into the
reservoir (DO NOT USE ALCOHOL, TAP WATER, OR DE-IONIZED WATER).
C.
Stop pouring when the water reaches the strainer visible at the bottom of the water fill
opening.
D.
Proceed with normal operations; always check the water level before starting.
5-2.
Maintenance of the Water Filter
The NORM-O-TEMP
®
circulating system includes a water filter designed to clear the line of any
particulate matter as the water is pumped through the system. The water filter assembly is a
plastic T-shaped fitting that intersects the hose from the water flow indicator to the reservoir. A
stainless steel wire-mesh screen is located inside the clear plastic cap of the water filter assembly.
Quarterly, or more often if deemed necessary, the water filter should be disassembled and cleaned.
To do so, the reservoir must first be drained and the three sided enclosure panel removed. The cap
of the water filter assembly is unscrewed. The wire mesh and the plastic cap are cleaned. The
parts are then reassembled.
After draining the reservoir as described in Section 5-1.4:
A.
Disconnect the unit from its power source.
CAUTION
•
Use distilled water only. Do Not Use De-Ionized water.
•
Do not
use alcohol. Alcohol may cause blanket/pad and unit deterioration.
•
Do not
overfill. Overfilling may result in overflow when the water in the blanket/pad drains
back into the system when the system is turned off.
CAUTION
•
Always drain the NORM-O-TEMP
®
system to a sanitary drain because
bacteria
may be present
in the unit’s water supply.
Содержание Norm-o-Temp 111Z
Страница 13: ...Operation Manual NORM O TEMP Model 111Z Page 13 of 50 Figure 1 NORM O TEMP unit Front View...
Страница 15: ...Operation Manual NORM O TEMP Model 111Z Page 15 of 50 Figure 2 NORM O TEMP unit Left Side View...
Страница 17: ...Operation Manual NORM O TEMP Model 111Z Page 17 of 50 Figure 3 NORM O TEMP unit Rear View...
Страница 19: ...Operation Manual NORM O TEMP Model 111Z Page 19 of 50 Figure 4 NORM O TEMP unit Right Side View...
Страница 21: ...Operation Manual NORM O TEMP Model 111Z Page 21 of 50 Figure 5 NORM O TEMP unit Model 111Z Top View...
Страница 49: ...NORM O TEMP Model 111Z Page 49 of 50 THIS PAGE INTENTIONALLY LEFT BLANK...