Electrical System
Electrical Characteristics:
100V, 50/60Hz, 8.2 Amps
Heater Power:
800 watts
Time to Heat:
23°C ± 2°C to 37°C in approximately 7
minutes at ambient temperature.
Power Cord:
3 Conductor, detachable, 16/18
AWG, hospital grade plug
Leakage Current:
Under 300
µ
a
Fuse
10 Amp in IEC 320 connector
Mains Supply Isolation:
Two-Pole Mains Switch
Warranty
1 year parts. Parts and labor if returned to
factory. Additional Warranty Available
Safety System
Maximum High Control Setting
42°C / 107.6°F
Primary Temperature Limit:
43.5°C ± 0.6°C (110.0°F ± 1.08°F)
Secondary Temperature Limit:
44.5°C ± 0.6°C (112.0°F ± 1.08°F)
Independent Mechanical Temperature Limit:
46.0°C ± 0.6°C (115.0°F ± 1.08°F)
Environmental Conditions
(during storage and transportation)
Ambient Temperature
(transportation and storage):
-40°C –
+
50°C (-40°F –
+
122°F)
Humidity (transportation and storage):
20% – 95%
UL Electrical Classification
Equipment is Class I.
Equipment is Type BF.
CERTIFICATIONS
Figure 6. NORM-O-TEMP
®
, Model 111Z Features
Medical Electrical Equipment
In Accordance With:
UL 60601-1 2
nd
Ed.
IEC 60601-1-2 3
rd
Ed.
ASTM F-2196-2
Содержание Norm-o-Temp 111Z
Страница 13: ...Operation Manual NORM O TEMP Model 111Z Page 13 of 50 Figure 1 NORM O TEMP unit Front View...
Страница 15: ...Operation Manual NORM O TEMP Model 111Z Page 15 of 50 Figure 2 NORM O TEMP unit Left Side View...
Страница 17: ...Operation Manual NORM O TEMP Model 111Z Page 17 of 50 Figure 3 NORM O TEMP unit Rear View...
Страница 19: ...Operation Manual NORM O TEMP Model 111Z Page 19 of 50 Figure 4 NORM O TEMP unit Right Side View...
Страница 21: ...Operation Manual NORM O TEMP Model 111Z Page 21 of 50 Figure 5 NORM O TEMP unit Model 111Z Top View...
Страница 49: ...NORM O TEMP Model 111Z Page 49 of 50 THIS PAGE INTENTIONALLY LEFT BLANK...