Operation Manual
NORM-O-TEMP
®
, Model 111Z
Page 35 of 50
Section 5.
General Maintenance of the NORM-O-TEMP
®
System
5-0.
Introduction
This section describes the general requirements maintenance personnel should complete on a
regular basis so that the NORM-O-TEMP
®
system continues to operate within the manufacturer’s
specifications.
These maintenance requirements include changing and replenishing the water in the reservoir,
and cleaning the water filter, the exterior of the unit, the back vent, and the hyperthermia
blanket/pad(s). Figure 7 provides a convenient checklist to record the required maintenance
history of the unit.
The description of each requirement includes the tools, the procedures, and the precautions
necessary to complete the requirements. Maintenance, repair and/or service of NORM-O-TEMP
®
by qualified maintenance/technical personnel will not void the warranty of the unit.
The following maintenance procedures should be performed quarterly at a minimum unless
otherwise specified below:
A.
Draining/Replenishing the Reservoir (perform monthly at a minimum)
B.
Required Preventative Maintenance Checklist
C.
Cleaning the Water Filter
D.
Internal Cleaning of Water System
E.
Cleaning Vents and Exterior of Unit (perform as necessary, minimum of quarterly)
F.
Cleaning Hyper-Hypothermia Blanket/pad(s)
Required Preventative Maintenance Checklist
WARNING
•
Before
returning the NORM-O-TEMP
®
to patient use
after
repairs, the FIRST TIME SET-
UP/SYSTEM TEST ROUTINE in Section 3-2 must
always
be performed.
Improper repair and
inadequate maintenance can result in damage to the NORM-O-TEMP
®
system and patient
injury.
Содержание Norm-o-Temp 111Z
Страница 13: ...Operation Manual NORM O TEMP Model 111Z Page 13 of 50 Figure 1 NORM O TEMP unit Front View...
Страница 15: ...Operation Manual NORM O TEMP Model 111Z Page 15 of 50 Figure 2 NORM O TEMP unit Left Side View...
Страница 17: ...Operation Manual NORM O TEMP Model 111Z Page 17 of 50 Figure 3 NORM O TEMP unit Rear View...
Страница 19: ...Operation Manual NORM O TEMP Model 111Z Page 19 of 50 Figure 4 NORM O TEMP unit Right Side View...
Страница 21: ...Operation Manual NORM O TEMP Model 111Z Page 21 of 50 Figure 5 NORM O TEMP unit Model 111Z Top View...
Страница 49: ...NORM O TEMP Model 111Z Page 49 of 50 THIS PAGE INTENTIONALLY LEFT BLANK...