5-1.2.
Internal Cleaning and Disinfection Using Gigasept FF / Dry Storage Procedure
For Circulating Water Units Utilizing CE Approved Gigasept FF (Product of Schulke &
Mayr) (Contact your local Cleaning Supplier). Gigasept FF Is Not Available In the United
States.
The following procedure disinfects the fluid circuit in these products. The required
tools/supplies are appropriate drain hose(s) and/or hose connector(s), Gigasept FF,
distilled water, and the appropriate AC electrical power
Procedure:
A.
Drain the water from the reservoir as described in Section 5-1.4.
B.
Turn unit off (“O” position).
C.
Fill the reservoir with the appropriate amount of Gigasept FF per chart below .
D.
Fill the reservoir with distilled water.
E.
Turn the unit on and circulate per the Chemical Cleaning Circulation Chart (above).
F.
Drain the unit as instructed in Step A.
G.
Rinse the unit with Distilled Water three (3) times as described in Steps D-F. Be sure to
omit Gigasept FF.
H.
After the third rinse, drain the unit and fill the reservoir with distilled water and
circulate.
I.
Check the water with PH strips or other appropriate test method for Gigasept FF
detection. If Gigasept FF is detected, repeat Steps F-H.
J.
If unit is to be
returned to service, replenish reservoir(s) with the appropriate volume of
distilled water and continue procedure at Step N.
K.
If
unit is being
placed in dry storage, continue with procedure.
L.
When all fluid has been removed from the unit, disconnect the drain hose(s) and wipe
unit clean.
M.
Unit is now ready for storage.
N.
Document unit maintenance.
UNIT
Gigasept FF
WATER
NORM-O-TEMP
®
, 111Z
168 Milliliters
5.4 Liters
Содержание Norm-o-Temp 111Z
Страница 13: ...Operation Manual NORM O TEMP Model 111Z Page 13 of 50 Figure 1 NORM O TEMP unit Front View...
Страница 15: ...Operation Manual NORM O TEMP Model 111Z Page 15 of 50 Figure 2 NORM O TEMP unit Left Side View...
Страница 17: ...Operation Manual NORM O TEMP Model 111Z Page 17 of 50 Figure 3 NORM O TEMP unit Rear View...
Страница 19: ...Operation Manual NORM O TEMP Model 111Z Page 19 of 50 Figure 4 NORM O TEMP unit Right Side View...
Страница 21: ...Operation Manual NORM O TEMP Model 111Z Page 21 of 50 Figure 5 NORM O TEMP unit Model 111Z Top View...
Страница 49: ...NORM O TEMP Model 111Z Page 49 of 50 THIS PAGE INTENTIONALLY LEFT BLANK...