Installation and Assembly
User’s Manual
4-9
.
Figure 4-8.
Double Limb Patient Circuit
Figure 4-9.
Close-up of Exhalation Valve Tube and Proximal Pressure Tube
Figure 4-9
shows details of the connections of the proximal pressure tube (
Figure 4-8
, item 6) and
the exhalation valve tube (
Figure 4-8
, item 8).
To connect a single limb circuit without an exhalation valve (NIV only):
refer to
Figure 4-10
.
1. Inspect the components of the patient circuit for any signs of damage, such as cracks (which
might cause leakage). Do not use damaged components to assemble the patient circuit.
2. Install the bacteria filter (item 1) on the TO PATIENT
outlet port, as shown.
3. Attach one end of the short circuit tubing (item 4) to the filter (item 1)
4. Attach the other end of the circuit tubing (item 4) to the inlet port of the humidifier (item 2).
Note:
Although shown here, the humidifier (item 2), water traps (item 3), and their connecting tubes are not
included with the patient circuit or ventilator. Contact your supplier for more information.
See also
Figure 4-9
on page 4-9.
1
2
11
4
10
9
8
6
5
7
3
13
12
14
Proximal Pressure
Tube
Exhalation
Valve Tube
Inhalation Port
Содержание Puritan Bennett 560
Страница 1: ...User s Manual Puritan Bennett 560 Ventilator TM...
Страница 6: ...This page intentionally blank iv User s Manual...
Страница 10: ...viii User s Manual This page is intentionally blank...
Страница 14: ...4 User s Manual This page is intentionally blank...
Страница 32: ...Safety Information 1 18 User s Manual Figure 1 4 Location of Labels Bottom View 7 13 1 4 3...
Страница 98: ...5 10 User s Manual This page is intentionally blank...
Страница 128: ...C 2 User s Manual This page is intentionally blank...
Страница 130: ...Unpacking and Preparation D 2 User s Manual Figure D 1 Puritan Bennett 560 Ventilator Figure D 2 Dual Bag...
Страница 134: ...E 4 User s Manual This page is intentionally blank...
Страница 148: ...Index Index 6 User s Manual This page intentionally blank...
Страница 149: ......