E
Alarms Tests
Before connecting the ventilator to the patient, perform the following tests to ensure the ventilator‘s alarms
are working properly.
Warning
Do not perform ventilator alarm tests while the patient is connected to the ventilator. Provide
the patient with an alternate means of ventilation before conducting these tests.
If the ventilator fails any alarm test or if you cannot complete these tests, see the
Troubleshooting section (refer to chapter
3, “Alarms and Troubleshooting”
) of this manual or
call your equipment supplier or Covidien (refer to section
8.3, “Service Assistance”
)
.
The setting of the Min PIP alarm must be adjusted for the patient, but must also be set high
enough to allow the PATIENT DISCONNECTION alarm to trigger properly. Perform the Low
Pressure Test (refer to section
E.1, “Low Pressure Test”
) to ensure the Min PIP alarm is properly
set.
E.1
Low Pressure Test
Warning
The setting of the Min PIP alarm must be adjusted for the patient, but must also be set high
enough to allow the PATIENT DISCONNECTION alarm to trigger properly. Perform the following
test to ensure the Low PIP alarm is properly set.
1. Before proceeding, set the ventilation and alarm parameters specified by the patient’s clinician
and specify a single or dual circuit setup.
2. Press the
VENTILATION ON/OFF
key to start ventilation.
3. Keep the patient‘s end of the breathing circuit open and allow ventilation to continue.
4. Wait for (Apnea Time + 2 seconds; Apnea time is not always 5 seconds), then ensure that:
•
the High priority indicator (red colour) lights up
•
the “PATIENT DISCONNECTION” alarm is displayed
•
the audible alarm sounds
5. Press the
ALARM CONTROL
key once to silence the alarm.
6. Press and hold the
VENTILATION ON/OFF
key for three (3) seconds, then release it.
Press the
VENTILATION ON/OFF
key again to confirm stop.The ventilator will switch to
Standby mode and cancel the alarms.
Note:
Many of the functions of the ventilator are not accessible when the Locking key
is enabled. For
additional assistance contact your Clinician or equipment representative.
Most of these tests require that an approved patient circuit be connected to the ventilator. Ensure that
your patient circuit is properly connected prior to performing these tests.
Содержание Puritan Bennett 560
Страница 1: ...User s Manual Puritan Bennett 560 Ventilator TM...
Страница 6: ...This page intentionally blank iv User s Manual...
Страница 10: ...viii User s Manual This page is intentionally blank...
Страница 14: ...4 User s Manual This page is intentionally blank...
Страница 32: ...Safety Information 1 18 User s Manual Figure 1 4 Location of Labels Bottom View 7 13 1 4 3...
Страница 98: ...5 10 User s Manual This page is intentionally blank...
Страница 128: ...C 2 User s Manual This page is intentionally blank...
Страница 130: ...Unpacking and Preparation D 2 User s Manual Figure D 1 Puritan Bennett 560 Ventilator Figure D 2 Dual Bag...
Страница 134: ...E 4 User s Manual This page is intentionally blank...
Страница 148: ...Index Index 6 User s Manual This page intentionally blank...
Страница 149: ......