Alarms and Troubleshooting
3-18
User’s Manual
COOLING FAN
RESTART/SRVC
Operating speed of the cooling fan not
properly adjusted for the internal ambient
temperature of the device.
Restart ventilator to see if alarm clears. If
not, replace the ventilator and call your
customer service representative.
DC POWER
DISCONNECTION
12 – 30 VDC power supply cut off when
there is no AC (“mains”) power supply.
Cancel the alarm then check the supply
wiring and/or the effective availability of
voltage on the external source.
Ventilator’s current-limiting fuse blown.
Replace the ventilator and call your
customer service representative.
DEVICE FAULT3 IF
PERSISTS RESTART/SRVC
24 V supply failure.
Restart ventilator to see if alarm clears. If
not, replace the ventilator and call your
customer service representative.
DEVICE FAULT5 IF
PERSISTS RESTART/SRVC
Internal problem in the electrical power
supply.
Restart ventilator to see if alarm clears. If
not, replace the ventilator and call your
customer service representative.
DEVICE FAULT7 IF
PERSISTS RESTART/SRVC
Internal technical problem.
Restart ventilator to see if alarm clears. If
not, replace the ventilator and call your
customer service representative.
DEVICE FAULT9 IF
PERSISTS RESTART/SRVC
POST RAM Error. RAM Read/Write does not
match memory setting.
If patient has been disconnected, reconnect
patient to reset the fault.
If persists restart ventilator to see if alarm
clears. If not, replace the ventilator and call
your customer service representative.
DEVICE FAULT10 IF
PERSISTS RESTART/SRVC
POST FLASH Checksum Error. Startup
FLASH computed checksum does not match
memory setting.
Restart ventilator to see if alarm clears. If
not, replace the ventilator and call your
customer service representative.
DEVICE FAULT11 IF
PERSISTS RESTART/SRVC
POST EEPROM Error. Startup EEPROM does
not match memory setting.
Restart ventilator to see if alarm clears. If
not, replace the ventilator and call your
customer service representative.
DEVICE FAULT12 IF
PERSISTS RESTART/SRVC
POST Reference Voltage Error. 5V or 10V
reference voltage error.
Restart ventilator to see if alarm clears. If
not, replace the ventilator and call your
customer service representative.
DEVICE FAULT13 IF
PERSISTS RESTART/SRVC
Incorrect software version detected.
Restart ventilator to see if alarm clears. If
not, replace the ventilator and call your
customer service representative.
EMPTY BATTERY
Internal battery capacity is less than 10 min.
(or 3%) —battery operation overextended.
Reconnect the device to an AC power
outlet, connect it to an external DC power
source, or replace the battery.
Reminder: the internal battery can be
charged only when the ventilator
connected to an AC power supply.
Table 3-2.
Alarms and Corrective Actions (Continued)
Alarm Message or
Symptom
Possible Reason(s) For The Alarm Event
Corrective Action(s)
Содержание Puritan Bennett 560
Страница 1: ...User s Manual Puritan Bennett 560 Ventilator TM...
Страница 6: ...This page intentionally blank iv User s Manual...
Страница 10: ...viii User s Manual This page is intentionally blank...
Страница 14: ...4 User s Manual This page is intentionally blank...
Страница 32: ...Safety Information 1 18 User s Manual Figure 1 4 Location of Labels Bottom View 7 13 1 4 3...
Страница 98: ...5 10 User s Manual This page is intentionally blank...
Страница 128: ...C 2 User s Manual This page is intentionally blank...
Страница 130: ...Unpacking and Preparation D 2 User s Manual Figure D 1 Puritan Bennett 560 Ventilator Figure D 2 Dual Bag...
Страница 134: ...E 4 User s Manual This page is intentionally blank...
Страница 148: ...Index Index 6 User s Manual This page intentionally blank...
Страница 149: ......