A Specifications
A.1
Physical
A.2
Electrical
Table A-1.
Physical Description (Excluding Accessories)
Ventilator Weight
9.9 lb. (4.5 kg)
Ventilator Dimensions
9.25 in wide x 12.40 in deep x 6.0 in high
(235 mm wide x 315 mm deep x 154 mm high)
Connectors
Inspiratory limb connector: ISO 22 mm (OD) conical
Exhalation limb connector (on exhalation block): ISO 22 mm (ID)
conical
Oxygen inlet: Female Connector with valve
Device airway volume
2000 ml
Breathing circuit volume
•
Adult, dual limb
1150 ml
•
Pediatric, dual limb
670 ml
•
Adult, single limb
550 ml
•
Pediatric, single limb
300 ml
Air Inlet Filter
Dimensions: 70 mm long x 60 mm wide
Composition: Polypropylene fiber electrostatic filter material, which is
laminated onto polyurethane open-celled foam.
Efficiency: 99.999982% at 30 lpm (filtering microbes 3.3 μm)
Inspiratory Bacteria
Filter Requirement
Maximum allowable flow resistance: 4mbar at 60 lpm
Table A-2.
AC Electrical Supply
Voltage
Frequency
Consumption
100 VAC to 240 VAC
50 Hz / 60 Hz
180 VA max
12 VDC
NA
8.3 A
30 VDC
NA
3.3 A
Содержание Puritan Bennett 560
Страница 1: ...User s Manual Puritan Bennett 560 Ventilator TM...
Страница 6: ...This page intentionally blank iv User s Manual...
Страница 10: ...viii User s Manual This page is intentionally blank...
Страница 14: ...4 User s Manual This page is intentionally blank...
Страница 32: ...Safety Information 1 18 User s Manual Figure 1 4 Location of Labels Bottom View 7 13 1 4 3...
Страница 98: ...5 10 User s Manual This page is intentionally blank...
Страница 128: ...C 2 User s Manual This page is intentionally blank...
Страница 130: ...Unpacking and Preparation D 2 User s Manual Figure D 1 Puritan Bennett 560 Ventilator Figure D 2 Dual Bag...
Страница 134: ...E 4 User s Manual This page is intentionally blank...
Страница 148: ...Index Index 6 User s Manual This page intentionally blank...
Страница 149: ......