Alarms and Troubleshooting
3-16
User’s Manual
BATTERY FAULT1
RESTART/SRVC
Battery problem that prevents it from
operating.
Restart ventilator to see if alarm clears. If
not,
replace the ventilator and call your
customer service representative.
BATTERY FAULT2
RESTART/SRVC
Internal battery missing or not detected.
Restart ventilator to see if alarm clears. If
not, replace the ventilator and call your
customer service representative.
BUZZER FAULT1
RESTART/SRVC
Defective operation of the buzzers.
Consequence: no audible tone when an
alarm is activated.
Restart ventilator to see if alarm clears. If
not, replace the ventilator and call your
customer service representative.
BUZZER FAULT2
RESTART/SRVC
Internal technical problem that prevents the
very high priority “POWER SUPPLY LOSS”
alarm from triggering.
Ensure that the protective cover over the I/O
switch located on the rear of the device is
intact and functioning properly. This cover
helps prevent accidental pressing of the I/O
switch and stoppage of the ventilation.
Ensure that the device is stabilised.
Call your customer service representative.
BUZZER FAULT3
RESTART/SRVC
Internal technical problem that prevents the
battery from correctly charging.
Restart ventilator to see if alarm clears. If
not, replace the ventilator and call your
customer service representative.
BUZZER FAULT4
RESTART/SRVC
Internal technical problem that prevents the
battery warning buzzer from sounding
POWER SUPPLY LOSS alarm.
Connect the ventilator to AC power and
switch on the device using the on/off at the
rear of the ventilator (I/O). Leave to charge
for at least 15 minutes.
If persists restart ventilator to see if alarm
clears. If not, replace the ventilator and call
your customer service representative.
BUZZER LOW BATTERY
Buzzer battery is too low to sound POWER
SUPPLY LOSS alarm.
Connect the ventilator to AC power and
switch on the device using the on/off at the
rear of the ventilator (I/O). Leave to charge
for at least 15 minutes.
CALIBRATE FIO
2
An FIO
2
sensor is detected and has not
been calibrated.
Calibrate FIO
2
sensor.
CALIBRATION FAIL
Too large a difference between a calibration
point and its tolerance range.
Restart calibration.
There may be a leak in the circuit. Ensure an
approved circuit is in use (refer to circuit
documentation).
Incorrect circuit type selected in the
Preferences menu.
Verify the circuit selection in the Preferences
menu matches the circuit in use.
Exhalation block defective or not properly
aligned.
Reset alarm message and ensure all
connections are secure, verify circuit
integrity, and verify the exhalation block is
properly seated.
Defective exhalation flow sensor.
Have a qualified technician replace the
defective component(s) and call your
customer service representative.
CHECK BATTERY
CHARGE
Battery charging impossible.
Do not disconnect the ventilator from the
AC power supply. Ensure that the power
cable is installed according to the
instructions in chapter
4, “Installation and
Assembly”
, so that the power cable cannot
be involuntarily disconnected.
In the event the internal battery capacity is
low, use an alternate device to ventilate the
patient.
Call your customer service representative.
Table 3-2.
Alarms and Corrective Actions (Continued)
Alarm Message or
Symptom
Possible Reason(s) For The Alarm Event
Corrective Action(s)
Содержание Puritan Bennett 560
Страница 1: ...User s Manual Puritan Bennett 560 Ventilator TM...
Страница 6: ...This page intentionally blank iv User s Manual...
Страница 10: ...viii User s Manual This page is intentionally blank...
Страница 14: ...4 User s Manual This page is intentionally blank...
Страница 32: ...Safety Information 1 18 User s Manual Figure 1 4 Location of Labels Bottom View 7 13 1 4 3...
Страница 98: ...5 10 User s Manual This page is intentionally blank...
Страница 128: ...C 2 User s Manual This page is intentionally blank...
Страница 130: ...Unpacking and Preparation D 2 User s Manual Figure D 1 Puritan Bennett 560 Ventilator Figure D 2 Dual Bag...
Страница 134: ...E 4 User s Manual This page is intentionally blank...
Страница 148: ...Index Index 6 User s Manual This page intentionally blank...
Страница 149: ......