Internal Battery
User’s Manual
6-3
Figure 6-1.
Internal Battery Indicator
•
A loss of external supply alarm is activated.
If ventilation is
stopped
, the internal battery reserve capacity is displayed as a percentage of battery charge.
Refer to
Figure 6-2
.
Figure 6-2.
Battery Reserve Capacity as a Percentage
If the ventilator is
running
, the internal battery reserve is momentarily displayed as a percentage. Then, after
the ventilator calculates the battery time remaining (which takes about two minutes, depending on the power
consumption of the ventilator), the internal battery reserve is then displayed in hours and minutes (rounded to
the nearest fifteen minutes). Refer to
Figure 6-3
.
Figure 6-3.
Battery Reserve Capacity in Hours and Minutes
The “LOW BATTERY“ and “EMPTY BATTERY“ alarms (refer to chapter
3, “Alarms and Troubleshooting”
) are
triggered when the internal battery reserve is reduced.
Содержание Puritan Bennett 560
Страница 1: ...User s Manual Puritan Bennett 560 Ventilator TM...
Страница 6: ...This page intentionally blank iv User s Manual...
Страница 10: ...viii User s Manual This page is intentionally blank...
Страница 14: ...4 User s Manual This page is intentionally blank...
Страница 32: ...Safety Information 1 18 User s Manual Figure 1 4 Location of Labels Bottom View 7 13 1 4 3...
Страница 98: ...5 10 User s Manual This page is intentionally blank...
Страница 128: ...C 2 User s Manual This page is intentionally blank...
Страница 130: ...Unpacking and Preparation D 2 User s Manual Figure D 1 Puritan Bennett 560 Ventilator Figure D 2 Dual Bag...
Страница 134: ...E 4 User s Manual This page is intentionally blank...
Страница 148: ...Index Index 6 User s Manual This page intentionally blank...
Страница 149: ......