Alarms and Troubleshooting
3-12
User’s Manual
HIGH RATE
Rate measured greater than
Max Rtot set during three
consecutive breaths.
Alarm activation occurs:
•
After three consecutive breaths.
MP
Yes
No
HIGH VTE
Expired tidal volume greater than
Max VTE set during three
consecutive breaths (in double limb
setup).
Alarm activation occurs:
•
After three consecutive breaths.
MP
Yes
No
HIGH VTI
Inspired tidal volume greater than
Max VTI set during three consecutive
breaths in PSV, CPAP, P A/C, P SIMV,
and V SIMV modes.
Alarm activation occurs:
•
After three consecutive breaths.
HP Yes
No
INSP FLOW
RESTART/SRVC
Inspiratory flow is constant
(± 1 lpm) with normal turbine
temperature and speed conditions.
Contact your service representative
for assistance.
HP
Yes
No
INTENTIONAL VENT
STOP
Ventilation has been stopped
voluntarily by the caregiver or
patient.
HP
Yes
Yes
KEYPAD FAULT
RESTART/SRVC*
*IF PERSISTS
RESTART/SRVC
Keyboard key held down for more
than 45 seconds.
HP No
No
LOW BATTERY
Internal battery capacity < 30 min.
or 8%.
HP Yes
No
LOW FIO
2
The level of oxygen delivered by the
ventilator is below the Min FIO
2
level
set.
MP
Yes
No
LOW VTE
Expired tidal volume less than Min
VTE set during three consecutive
breaths (in double-limb setup).
Alarm activation occurs:
•
After three consecutive breaths.
MP
Yes
No
LOW VTI
Inspired tidal volume less than Min
VTI set during three consecutive
breaths in PSV, CPAP, P A/C, P SIMV
and V SIMV modes.
Alarm activation occurs:
•
After three consecutive breaths.
MP
Yes
No
NO
PROXIMAL LINE2*
*IF PERSISTS
RESTART/SRVC
Proximal pressure < 0.6 mbar for
100 ms during inspiration phase of
3
rd
breath cycle.
Ventilator response: Switch to
internal pressure sensor for pressure
measurement.
MP
Yes
No
Table 3-1.
Overview of Alarms (Continued)
Alarm Message
Cause/Ventilator Response
Priority
Audio
Paused
Avail.
Alarm
Paused
Avail.
Содержание Puritan Bennett 560
Страница 1: ...User s Manual Puritan Bennett 560 Ventilator TM...
Страница 6: ...This page intentionally blank iv User s Manual...
Страница 10: ...viii User s Manual This page is intentionally blank...
Страница 14: ...4 User s Manual This page is intentionally blank...
Страница 32: ...Safety Information 1 18 User s Manual Figure 1 4 Location of Labels Bottom View 7 13 1 4 3...
Страница 98: ...5 10 User s Manual This page is intentionally blank...
Страница 128: ...C 2 User s Manual This page is intentionally blank...
Страница 130: ...Unpacking and Preparation D 2 User s Manual Figure D 1 Puritan Bennett 560 Ventilator Figure D 2 Dual Bag...
Страница 134: ...E 4 User s Manual This page is intentionally blank...
Страница 148: ...Index Index 6 User s Manual This page intentionally blank...