Troubleshooting Guide
Operator's Manual
5-7
WARNING:
If, at any time, the patient is not responding to ventilation appropriately, the patient should be
taken off the ventilator immediately and connected to an alternate method of ventilation. Contact
your health care provider or physician immediately.
Note:
To ensure best battery performance, always plug the ventilator into an external power source when it is
available, even when the ventilator is not in use.
5.3
Troubleshooting Guide
Ventilation and alarm settings are determined by the physician’s prescription. Consult with the
physician regarding ventilation and alarm settings.
Note:
Clear alarm messages with the
Audio Paused/Reset
button.
Note:
Review the entire operator’s manual for full user instructions.
Note:
Note that the minute volume alarms are expiratory minute volume alarms when the on-airway flow sensor
is in use and they are inspiratory minute volume alarms when the on-airway flow sensor is not in use.
Note:
Backup Ventilation increases the respiratory rate in A/CMV and SIMV and provides pressure-controlled
breaths in SPONT mode. It may be caused by the violation of a Low Minute Volume alarm or Apnea alarm.
Resolve the alarm to resolve the Backup Ventilation.
Table 5-2.
Troubleshooting
Problem/area of concern
Probable cause
Resolution
Touch screen does not respond
to touch
Screen is locked by Autolock
function (set on Utility screen).
Touch the lock icon in the lower right corner of the
screen for 3 seconds to unlock the screen.
Using basic domain and the
parameters are changed to
displays rather than buttons.
Touch the
Basic Domain
button in the lower right corner
of the screen to toggle the domain to hospital (Hosp),
and then press
Accept
.
Touch screen is dark/blank
Power Save is on (Utility screen)
and the screen has “gone to sleep”
to save power.
Touch the screen or push any button to “wake up” the
screen.
Need to view all monitored
values
N/A
Touch any monitored value at lower edge of the screen.
Содержание Newport HT70
Страница 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Страница 8: ...Page Left Intentionally Blank viii Operator s Manual...
Страница 12: ...Page Left Intentionally Blank xii Operator s Manual...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Страница 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Страница 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Страница 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Страница 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Страница 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Страница 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Страница 142: ...Foldout Drawings B 2 Page left intentionally blank...
Страница 144: ...Foldout Drawings B 4 Page left intentionally blank...
Страница 146: ...Foldout Drawings B 6 Page left intentionally blank...
Страница 147: ......