Operating the Ventilator
5-6
Operator's Manual
2.
Ensure the ventilator, patient circuit, and accessories are assembled correctly, as described in Chapter
3.
Make sure the ventilator has passed the quick check procedure.
4.
Perform the circuit check. Resolve any issues. Reference
on page
5.
Set all parameters per physician’s prescription using manual adjustment or a custom or default preset.
Reference Chapter
for specifications on all settings.
a.
Select mode and breath type. Set all parameters on the Main screen and relevant parameters on
the More screen.
b.
Select safe/appropriate alarm limits on the Alarms screen.
c.
Ensure that the alarm loudness is set loud enough for the alarm to be heard under all circumstanc-
es.
6.
Place a test lung on the patient end of the breathing circuit, and press the
Start Ventilation
button on
the touch screen.
7.
Ensure that the ventilator starts operating appropriately.
Note:
Flow and volume are set and measured at the ambient temperature pressure saturated (ATPS) condition.
Note:
While ventilating a test lung, peak pressure for volume breaths and volume delivery for pressure breaths
will be different than they will be on the patient. Pressure Support breaths will not perform the same way
they do on patients. And PEEP may cause auto-triggering. These differences may cause nuisance alarms
during this step.
8.
When ready, remove the test lung and attach the patient connection of the breathing circuit to the
patient interface.
9.
Monitor the patient settings and check for appropriate alarm settings.
10.
Verify that the patient trigger icon lights each time the patient initiates a spontaneous breath. Re-
adjust sensitivity (Ptrig or Flow trig if using on-airway flow sensor) as necessary to ensure comfortable
triggering without auto-triggering. When using PEEP while ventilating a patient with an airway leak,
set NIV to on and adjust bias flow to stabilize the PEEP and eliminate auto-triggering at reasonable
trigger settings.
11.
Closely monitor the patient and ventilator to ensure appropriate oxygen delivery, and adequate oxy-
genation and ventilation.
WARNING:
Always ensure adequate monitoring is in place when ventilating patients.
Содержание Newport HT70
Страница 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Страница 8: ...Page Left Intentionally Blank viii Operator s Manual...
Страница 12: ...Page Left Intentionally Blank xii Operator s Manual...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Страница 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Страница 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Страница 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Страница 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Страница 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Страница 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Страница 142: ...Foldout Drawings B 2 Page left intentionally blank...
Страница 144: ...Foldout Drawings B 4 Page left intentionally blank...
Страница 146: ...Foldout Drawings B 6 Page left intentionally blank...
Страница 147: ......