Automatic Alarms
Operator's Manual
6-9
6.5.9
Device Alert Alarm—System Error
The Device Alert alarm is activated when the microprocessor detects a functional problem with
the ventilator. When this occurs, an alternate means of ventilation should be used. The ventilator
must be powered down by pressing the on/off button on the back of the unit.
If the cause of the device alert does not allow the ventilator to display the alarm message and the
device alert indicator to light, the ventilator will shut down, and the Shut Down Alert alarm will
activate.
WARNING:
If a device alert alarm occurs, immediately disconnect the patient from the ventilator and provide
an alternate method of ventilation until the cause of the alert has been determined and corrected.
6.5.10
Motor Fault Alarm
The Motor Fault alarm is activated when the microprocessor detects a functional problem with
the motor or motor control systems. When this occurs, the ventilator should be replaced and sent
in for service.
6.5.11
Shut Down Alert Alarm
The Shut Down Alert alarm occurs when the ventilator is turned off. A continuous audible alert
indicates the ventilator is no longer operating. The alert beeps will continue for at least 15 minutes
or until muted by pressing the
Audio Paused/Reset
button.
6.5.12
Internal Temperature Alarm
The Internal Temperature alarm indicates that the internal temperature has exceeded specifica-
tions. Environmental temperature during operation should not exceed 40°C (104°F). Connect the
ventilator to an external power source as soon as possible and take steps to make the environ-
ment cooler. Also check that the fan filter is clean.
6.5.13
O
2
Cylinder Low/Empty Alarms
This alarm indicates that the oxygen cylinder that was set up in the O
2
calculator data screen has
reached a low or empty level. The O
2
Cylinder Low alarm will sound when the time reaches 10
minutes. The O
2
Cylinder Empty alarm will sound when the estimated time reaches 5 minutes.
Содержание Newport HT70
Страница 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Страница 8: ...Page Left Intentionally Blank viii Operator s Manual...
Страница 12: ...Page Left Intentionally Blank xii Operator s Manual...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Страница 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Страница 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Страница 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Страница 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Страница 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Страница 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Страница 142: ...Foldout Drawings B 2 Page left intentionally blank...
Страница 144: ...Foldout Drawings B 4 Page left intentionally blank...
Страница 146: ...Foldout Drawings B 6 Page left intentionally blank...
Страница 147: ......