Specifications
9-6
Operator's Manual
Table 9-7.
User-Adjustable Alarms
Alarm
Alarm priority
Range/description
1
P (High Pressure)
High
5-99 cmH
2
O / 5-99 mbar
3
P (Low Pressure)
High
NIV off: 3-98 cmH
2
O / 3-98 mbar (limited by PEEP+3); two-breath delay
NIV on: 1-98 cmH
2
O / 1-98 mbar (limited by PEEP+1); three-breath
delay
1
Min Vol (High Insp./Exp. Minute
Volume)
High
NIV off: 1.1-50 L/min
NIV on: 1.1-80 L/min
3
Min Vol (Low Insp./ Exp. Minute
Volume)
High
NIV off: 0.01-49.0 L/min
NIV on: Off, 0.01-49.0 L/min
1
RR (High Respiratory Rate)
Med
Off, 30-100 b/min
Apnea
High
5-70 seconds
1
O
2
Med
Off, 24-100, only available when O
2
sensor is enabled
3
O
2
High
Off, 22-98, only available when O
2
sensor is enabled
1
VTE
High
OFF, 0.06-2.2 Liters
Backup Ventilation (BUV)
Med
Can be set to be activated by either the Low Minute Volume alarm or
the Apnea alarm or both via the More>Utilities>Custom Settings>BUV
settings. Functional in all modes.
Table 9-8.
Automatic Alarms
Alarm
Alarm priority
Description
Low Baseline Pressure
High
Paw <PEEP minus 2 cmH
2
O/mbar for 3 sec.
High Baseline Pressure
Med
Paw 5 above set PEEP at onset of a time-triggered breath.
Pressure Control Setting Not
Reached
High
P Peak <50% of PCV setting for two consecutive breaths.
Check Circuit
High
Circuit may be disconnected or proximal pressure line may be pinched
or blocked.
No External Power
Low
Loss of external power, automatic switchover to internal dual battery
system.
Occlusion
High
An occlusion or restriction in the circuit that interferes with exhalation.
Occlusion - Sustained
High
Occlusion continues for 10 sec or two breath periods, whichever is
shorter.
Device Alert
High
Ventilator malfunction, device alert LED lights red.
Shut Down Alert
High
Mute by pressing
Audio Paused/Reset
button.
Motor Fault
High
Hardware detected fault in the motor drive circuit has occurred.
Internal Temperature
Low
Internal temperature is >60°C.
Содержание Newport HT70
Страница 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Страница 8: ...Page Left Intentionally Blank viii Operator s Manual...
Страница 12: ...Page Left Intentionally Blank xii Operator s Manual...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Страница 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Страница 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Страница 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Страница 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Страница 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Страница 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Страница 142: ...Foldout Drawings B 2 Page left intentionally blank...
Страница 144: ...Foldout Drawings B 4 Page left intentionally blank...
Страница 146: ...Foldout Drawings B 6 Page left intentionally blank...
Страница 147: ......