Miscellaneous Specifications
Operator's Manual
9-11
9.11.2
Ventilator Breathing System (VBS) Resistances
9.11.3
Pneumatic Requirements (Optional Equipment)
WARNING:
Appropriate oxygen monitoring is required for patient safety.
Note:
Oxygen source gas must be medical-grade, 100% oxygen.
9.11.4
Manufacturer’s Declaration
The following tables contain the manufacturer’s declarations for the Newport™ HT70 Ventilator’s
electromagnetic emissions, electromagnetic immunity, and recommended separation distances
between the ventilator and portable and mobile RF communications equipment.
WARNING:
Portable and mobile RF communications equipment can affect the performance of the ventilator
system. Install and use this device according to the information contained in this manual.
WARNING:
The ventilator system should not be used adjacent to or stacked with other equipment, except as
may be specified elsewhere in this manual. If adjacent or stacked used is necessary, the ventilator
Table 9-15.
VBS Resistances
Inspiratory resistance
Expiratory resistance
0.32 kPa at 60 L/min
0.28 kPa at 60 L/min
0.20 kPa at 30 L/min
0.14 kPa at 30 L/min
0.03 kPa at 5 L/min
0.02 kPa at 5 L/min
Table 9-16.
Pneumatic Requirements
Air/oxygen entrainment mixer
Low flow oxygen reservoir
Oxygen
35-65 psig (2.4-4.5 Bar) full operating range, maximum accuracy
40-50 psig (2.8-3.4 Bar) accuracy ±0.08
0-10 L/min
Air
Atmospheric pressure
Atmospheric pressure
FiO
2
Control
Adjusted continuously from 0.21 to 1.00
FiO
2
, indirectly adjusted from 0.21 up
to 1.00 via oxygen flow (L/min)
Содержание Newport HT70
Страница 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Страница 8: ...Page Left Intentionally Blank viii Operator s Manual...
Страница 12: ...Page Left Intentionally Blank xii Operator s Manual...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Страница 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Страница 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Страница 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Страница 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Страница 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Страница 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Страница 142: ...Foldout Drawings B 2 Page left intentionally blank...
Страница 144: ...Foldout Drawings B 4 Page left intentionally blank...
Страница 146: ...Foldout Drawings B 6 Page left intentionally blank...
Страница 147: ......