Attaching a Patient Circuit
Operator's Manual
3-13
Figure 3-12.
Patient Circuit with HME (Completed Setup)
12.
Perform the circuit check test. If the circuit includes an end cap, retain it for use in the circuit check test.
Reference
on page
3.6.3
Using the On-Airway Flow Sensor
The Newport flow sensor is a disposable, single-patient use on-airway flow sensor that can be
used for pediatric to adult patients. Use the 15/22 mm adapter supplied with the on-airway flow
sensor to attach to the patient interface as needed. Use the circuit clip supplied with the on-
airway flow sensor to secure the on-airway flow sensor lines to the main breathing circuit tubing.
Set up the patient breathing circuit as described in
usage (with or without humidifier or HME). Plug the on-airway flow sensor connector into the port
on the front panel of the ventilator. Attach the on-airway flow sensor to the patient end of the
breathing circuit with the blue tubing toward the patient.
When it is connected, the ventilator will recognize the on-airway flow sensor and enable these
added features:
•
Flow trigger
•
Exhaled volume monitoring
•
High Tidal Volume alarm
•
High and Low Expiratory Minute Volume alarms
Содержание Newport HT70
Страница 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Страница 8: ...Page Left Intentionally Blank viii Operator s Manual...
Страница 12: ...Page Left Intentionally Blank xii Operator s Manual...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Страница 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Страница 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Страница 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Страница 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Страница 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Страница 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Страница 142: ...Foldout Drawings B 2 Page left intentionally blank...
Страница 144: ...Foldout Drawings B 4 Page left intentionally blank...
Страница 146: ...Foldout Drawings B 6 Page left intentionally blank...
Страница 147: ......