Battery Operation
7-2
Operator's Manual
Figure 7-1.
Power Pac
When the ventilator is used for transport applications, ensure that the Power Pac battery pack is
fully charged prior to use. It is highly recommended to carry an extra, fully charged Power Pac
during transport or outdoor applications.
The Power Pac can be charged independently from the ventilator and has an LED on the bottom
edge to show charge condition. To view the charge condition, push the button (green = approx-
imately 90% or higher charge level, amber = charge not completed, red = battery depleted).
Always attach the Power Pac to the ventilator and turn it on to verify the actual charge level per-
centage (shown in the message display).
*Standard Settings:
Fully charged, new battery in good condition. Power Save on. Peak pres-
sures below 30 cmH
2
O with these settings:
•
Mode = A/CMV
•
Respiratory rate = 15
•
Tidal volume = 500 mL
•
Inspiratory time = 1.0 seconds
•
PEEP = 0
7.1.2
Backup Battery
The secondary backup battery will provide a minimum of 30 minutes of operation. The ventilator
will automatically switch to the backup battery whenever the Power Pac battery pack is removed
or when the Power Pac power is low and the Switching to Backup Battery alarm occurs.
7.2
Conditions that Affect Battery Use Time
Each of the items listed below will affect the amount of time that the internal dual battery system
will last:
e
l
t
Ex
a
rn
w
r
Po
e
l
M
au
na
fa
n
n
to
l
i
gh
r
t
s
B
i
se
n
ve
ec
D
i
e
A
tr
l
Содержание Newport HT70
Страница 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Страница 8: ...Page Left Intentionally Blank viii Operator s Manual...
Страница 12: ...Page Left Intentionally Blank xii Operator s Manual...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Страница 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Страница 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Страница 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Страница 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Страница 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Страница 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Страница 142: ...Foldout Drawings B 2 Page left intentionally blank...
Страница 144: ...Foldout Drawings B 4 Page left intentionally blank...
Страница 146: ...Foldout Drawings B 6 Page left intentionally blank...
Страница 147: ......