User-Adjustable Alarms
Operator's Manual
6-3
When the violation is no longer in effect, the alarm message latches (remains steadily visible) until
it is reset by pressing the
Audio Paused/Reset
button.
6.2.1
Audio Paused/Reset Button
Press the
Audio Paused/Reset
button to mute the audible alarm for 1 minute (60 sec). Once an
alarm condition is corrected, press this button to clear (reset) the alarm message. Press repeatedly
to clear multiple messages. Press and hold for 3 seconds to clear all alarm messages at once.
WARNING:
Failure to identify and correct alarm violations may result in patient injury.
6.2.2
Audio Pause LED
Located next to the
Audio Paused/Reset
button, the audio pause LED remains lit during the 1
minute audio pause period.
6.3
User-Adjustable Alarms
+ only available on the HT70PM when the on-airway flow sensor is in use
The HT70M model ventilator monitors the volume of gas output from the ventilator; therefore,
the High and Low Inspiratory Minute Volume alarms respond to changes in delivered, not
exhaled, volumes. For this model, use the High Inspiratory Minute Volume Alarm (
1
Min Vol) to
detect leaks or disconnects while using pressure-controlled ventilation, and use the Low Pressure
Alarm (
3
P) to detect circuit leaks or disconnect while using volume-controlled ventilation.
For the HT70PM model, if the on-airway flow sensor is in use, the sensor measures exhaled vol-
umes. So in this case, the ventilator will automatically display exhaled tidal volume and exhaled
minute volume, and the alarms will respond accordingly. That is, when the on-airway flow sensor
is in use, the Low Minute Volume alarm will respond to leaks and disconnects when minute
volume leaving the patient drops to the alarm setting. If exhaled volumes are important to the
care of the patient, Covidien recommends the HT70PM model with the on-airway flow sensor. To
verify exhaled volumes when not using the flow sensor, use a separate exhaled volume monitor.
Table 6-1.
User-Adjustable Alarms
1
P (High Pressure)
1
Min Vol (High Inspiratory or Exp Minute Volume)
3
P (Low Pressure)
3
Min Vol (Low Inspiratory or Exp Minute Volume)
1
RR (High Respiratory Rate)
1
O
2
(High O
2
)
Apnea (time adjustment)
3
O
2
(Low O
2
)
1
VTE (High Tidal
Содержание Newport HT70
Страница 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Страница 8: ...Page Left Intentionally Blank viii Operator s Manual...
Страница 12: ...Page Left Intentionally Blank xii Operator s Manual...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Страница 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Страница 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Страница 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Страница 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Страница 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Страница 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Страница 142: ...Foldout Drawings B 2 Page left intentionally blank...
Страница 144: ...Foldout Drawings B 4 Page left intentionally blank...
Страница 146: ...Foldout Drawings B 6 Page left intentionally blank...
Страница 147: ......