10
6.8 CLEANING AND STORAGE
1.
The ALLY UPS system, including the main unit, fixed arm, flexible arm, handle and foot pedal should be
cleaned after each use. Only use Isopropyl Alcohol 70 percent (70%) or higher. Wipe all surfaces free of
debris. Do not spray flexible arm. Ensure flexible arm is thoroughly dry before powering on.
2.
After each use clean the rail clamp; clean and disinfect using a quaternary ammonium disinfecting/
cleaning solution, read the cleaning product’s directions and follow the instructions on the label for
recommendation for achieving low-level disinfection; use caution around the knob where fluid migration
may occur. Wipe device with a clean, dry cloth; make certain the product is dry prior to reinstalling to
avoid damage.
3.
The foot pedal cover should be used to lift, carry or reposition the foot pedal.
4.
Store this device in a clean, dry and well-ventilated environment.
SECTION 7 GUIDANCE AND MANUFACTURER’S DECLARATION
– ELECTROMAGNETIC EMISSIONS
(for all Medical Equipment and Medical Equipment Systems)
The ALLY UPS is intended for use in the electromagnetic environment specified below. The customer or the user
of the ALLY UPS should assure that it is used in such an environment.
EMISSIONS TEST
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT GUIDANCE
RF emissions CISPR 11
Group 1
The ALLY UPS uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11
Class A
The ALLY UPS is suitable for use in all establishments
other than domestic and those directly connected to
Harmonic emissions IEC 61000-3-2
Class A
the public low voltage power supply network that
supplies buildings used for domestic purposes.
Voltage fluctuations/flicker emissions
Complies
IEC 61000-3-3
CAUTION:
Damage may result if the rail clamp is cleaned with bleach.
!
CAUTION:
Care should be taken during storage and transportation to avoid drops, falls, mechanical
stress, and mechanical shock to the device.
!
CAUTION:
Damage will result if the flexible arm is cleaned with bleach products.
!
WARNING: DO NOT
store the device in direct sunlight, at high temperatures or high humidity.
!
WARNING: DO NOT
store this device in the shipping box.