Notes:
Page 62 of 107
SmartLF Gx Service Manual
5.
Run the PlotWorks.exe (the scanner must be attached and
seen by the PC to open)
The PlotWorks program was designed for use in production, so
many features are not to be used, please observe this, as incorrect
use of this program may corrupt the main board.
6.
Before any alignment, 'Un Normalise' the scanner.
7.
Open the PC scope function. Then select the camera requiring
alignment (Start with Camera 1 as all cameras will need
alignment which ever camera was replaced).
For initial firmware download in production,
DO NOT USE
Used to configure scanner type, using this feature will not
harm the scanner, set to T or non-T
before
setting to Gx25
if necessary. Just clicking the 'configure T model' or '….
Non T… ' will by default return the scanner to a 42" model.