10.2
Calibration
The calibration must be performed at the same wavelengths at which the sample will be measured. Insert a cuvette
containing the blank solution into the sample chamber and close the instrument lid. Touch the
icon. This sets
the instrument to zero absorbance and 100% transmittance.
10.3
Sample Measurement
It is not possible to measure a sample before the instrument has been calibrated at the selected wavelengths.Once
the calibration has been performed the
icon becomes active and a sample can be measured. Remove the
cuvette containing the blank solution and place a cuvette containing the sample to be measured in the sample
holder. Close the instrument lid and touch the
icon. Once the measurement is completed the result will
be shown on the screen. Touch
to measure subsequent samples can be measured in the same way. If
any of the wavelengths are adjusted between sample measurements then the instrument must be calibrated again
before more samples can be measured.
44