52
12.3.1.2 Automatic 8 Cell Turret - Automatic Mode
Set to automatic mode. Select the measurement mode you require (photometrics has been used as an example
only), the automatic 8 cell turret will automatically move to position 0.
Set up the required measurement parameters. Insert the cuvette containing the blank solution into turret
position 0 followed by the cuvettes containing the samples into turret positions 1 to 7 and close the instrument lid.
Touch and then touch
twice at position 1. This will activate the sample icon
.
Touch
once at each of the positions you have inserted samples into. This will activate the sample icon
.
Touch
to apply changes and return to method set up or touch
to return to method set up without
saving any changes.