8.1.2
Selecting Number of Replicates
To select the number of repeat measurements of a calibration standard touch
. Touch the circle
adjacent
to the required unit of replicates to apply and return to the method set up.
If 2 or more replicates are selected,
becomes active. Automatic replicates will read the
same sample for the selected number of replicates. If individual sample replicates are being used, do not select the
automatic replicates option.
8.1.3
Selecting Concentration Units
The units of concentration can be selected from several options. Touch
to select from the menu. Touch the
circle
adjacent to the required unit of concentration. The selected unit will be displayed against the final
concentration result.
Once the method parameters have been entered touch
to start measuring the calibration standards. Touch
to return to the home screen.