2201R Mobile Tester SCPI Reference Guide
Chapter 7 Format commands
47090/323
Copyright © 2016 Aeroflex Limited. All rights reserved.
80
Introduction
The FORMat subsystem sets and queries settings concerning the data output in remote mode.
FORMat subsystem
FORMat:MRESult:STYPe
Syntax
FORMat:MRESult:STYPe <PredefExp>
Parameters
PredefExp is one of the following predefined expressions: STB|SIGNalling|MEASuring|OPERation|QUEStionable|ALL.
Default is that no additional information is provided with the measurement result values.
Description
Specifies the type of additional information to be returned with any measurement result obtained by a FETCh
command. The meaning of the settings is as follows:
STB delivers the current contents of the Service register.
SIGNalling stands for the current contents of the signaling operation condition register.
MEASuring means that the current contents of the measuring operation condition register are delivered back.
OPERation delivers the current contents of the general operation condition register.
QUEStionable stands for the current contents of the general questionable status condition register.
ALL delivers the current contents of the eight most important registers. The order is as follows:
1. Service register,
2. event status register,
3. general operation condition register,
4. signaling operation condition register,
5. measuring operation condition register,
6. general questionable status condition register,
7. RF questionable status condition register, and
8. synchronization questionable status condition register.
The contents of every single register are returned as an integer; the single values are separated by commas.
Query
There is no query form of this command available.
Example
:FORMat:MRESult:HEADer ON
:FORMat:MRESult:STYPe ALL
:MEASure:RFTX:PRMS
:FETCh:RFTX:PRMS
The first command switches the transmission of the additional information on, the second command specifies that
all current contents of the eight most important registers shall be returned.
The third command starts a continuous measurement. The last command finally delivers the latest measurement
result value plus the current contents of the main registers. The string delivered back:
"0,128,256,8,1,0,0,0,4.63".