2201R Mobile Tester SCPI Reference Guide
Chapter 10 Measurement commands
47090/323
Copyright © 2016 Aeroflex Limited. All rights reserved.
251
:MEASure:EGPRs[:CONTinuous]:RFTX:FREQuency
Syntax
:MEASure:EGPRs[:CONTinuous]:RFTX:FREQuency
Parameters
There are no parameters.
Description
Starts a continuous measurement of the mobile frequency error in GSM mode. To read out the latest measurement
result, use the :FETCh:EGPRs:RFTX:FREQuency command.
Query
The query form of this command starts the measurement and returns a measurement result value as soon as the first
measurement has been completed. The string delivered back contains one floating point real number with the
physical dimension of Hertz.
Example
:MEASure:EGPRs:CONTinuous:RFTX:FREQuency?
The string returned in this example is: "-31.92".
:MEASure:EGPRs[:CONTinuous]:RFTX:LENGth
Syntax
:MEASure:EGPRs[:CONTinuous]:RFTX:LENGth
Parameters
There are no parameters.
Description
Starts a continuous measurement of the burst length in GSM mode. To read out the latest measurement result, use
the :FETCh:EGPRs:RFTX:LENGth command.
Query
The query form of this command starts the measurement and returns a measurement result value as soon as the first
measurement has been completed. The string delivered back contains one floating point real number with the
physical dimension of microseconds.
Example
:MEASure:EGPRs:CONTinuous:RFTX:LENGth?
The string returned in this example is: "557.0".